IP address
IP
address
T
Use of USB main control port connector on front panel
Type-A connector on the front panel is the connector of USB master control port. In S1465 series
signal generator, this port is used to connect the flash disk of USB interface to upgrade the
instrument TSR software. The signal generator can also be controlled by connecting USB keyboard
and mouse.
ON
5.2.1.1 Connection
S1465 signal generator and external controller (computer) are connected to LAN using network
cable, as shown in Fig.5.4:
Fig.5.4 LAN Interface Connection
5.2.1.2 Interface configuration
When the signal generator is remotely controlled via LAN, the physical network connection shall
be guaranteed to be smooth. As the signal generator does not support DHCP, domain name access
and wide-area network connection, the network programmable setting is relatively simple. In system
menu, “IP address”, “subnet mask” and “default gateway” are set to the subnet where the master
controller is located.
CAUTION
Ensure that the signal generator is physically connected properly by 10Base-T LAN or
100Base-T LAN cable
As the signal generator only supports the construction of a single LAN control system and the
setting of static IP address, rather than DHCP and the access to the host via DNS and domain name
server, users do not need to modify the subnet mask. In the instrument, it is fixedly set to: 255. 255.
255. 0.