5-37
© Sage Instruments 2000
9100-2930-02 01/31/00
Sage Instruments 930i Communications Test Set
5-11.1.1 Noise Threshold
To change the noise threshold:
1. Press K1. The display reads:
SET THRESHOLD: 67dBrnC [30-102]
K1
K2
K3
K4
2. Use the numeric keypad to enter a new threshold.
• A valid threshold is between
30
and
102 dBrnC
.
3. Press any soft key (K1-K4) to return to the second SET-UP display.
5-11.1.2 Spread
To change the spread:
1. Press K2 under
SPRD: N dB
from the
THRESH
display. The display reads:
SPRD: +8dB +6dB +4dB +2dB
K1
K2
K3
K4
8 dB
spread is not available with thresholds above
96 dBm
.
2. Press the appropriate soft key (K1-K4). You will be returned to the second Set-Up display.
5-11.1.3 Measurements Per Second
To enter the desired measurements per second:
1. Press K3 under
8 M/S
. The display reads:
MEASUREMENTS PER SECOND: 7 [1-99]
K1
K2
K3
K4
2. Use the numeric keypad to enter a new number of measurements per second.
• A valid number of measurements per second is between
1
and
99
.
Option Menu #:
11
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
Содержание 930i
Страница 11: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com ...
Страница 19: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com ...
Страница 43: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com ...
Страница 203: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com ...
Страница 235: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com ...