Sage Instruments 930i Communications Test Set
5
-10 OPTION MENU #:10 FREQUENCY SWEEP
OPTION MENU #:10, FREQUENCY SWEEP is a standard feature that allows the 930i to sweep up or
sweep down in frequency, either one time or continuously.
To sweep up or down in frequency:
1. Press the Option Menu function key.
2. Use the Up/Down Arrow keys to scroll through the option menus, OR select 10 using the
numeric keypad and them press the Option Menu function key. The display reads:
OPTION MENU #: 10 FREQUENCY SWEEP
K1
K2
K3
K4
3. Press ENT or any soft key (K1-K4) to activate your selection. The display reads:
FREQUENCY SWEEP: SET-UP SWEEP EXIT
K1
K2
K3
K4
5-10.1 Setup
To change the default parameters of the frequency sweep:
Press K2 under
SET-UP
. The
SELECT
display reads:
SELECT: BOUNDS STEP TIME/LEVEL EXIT
K1
K2
K3
K4
5-10.1.1 Bounds
BOUNDS are the beginning and ending frequencies between which the 930i conducts its sweep.
To change the starting or ending frequency parameters:
1. Press K1 under
BOUNDS
. The display reads:
STARTING FREQUENCY: 304 Hz
K1
K2
K3
K4
Option Menu #:
10
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
Содержание 930i
Страница 11: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com ...
Страница 19: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com ...
Страница 43: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com ...
Страница 203: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com ...
Страница 235: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com ...