© 2004 SAGE INSTRUMENTS
9100-0925-01 V1.2
16
T
ESTLINE
T
IMEOUT
When operating in a continuous test director mode you must manually stop Echo Sounder
and PVIT Send. See Remote Control of the 925VST on page 38 for more information.
For PVIT Measure in both test director mode and responder mode, PVIT Send in
responder mode, and Echo Generator (responder mode) you can use this setting to specify
how long the 925VST will run a test. When the maximum run-time has expired, the test
display switches to the results display. At the Testline Timeout screen, enter the number of
seconds before the 925VST automatically ends a test. The maximum run-time for
continuous mode director tests is 999 seconds; the default is 300 seconds.
SMOS T
ESTLINE
TPT
If the network under test is exceptionally slow in establishing and end-to-end talk path
between the director and responder units, you can adjust the duration of the test progress
tone (answer tone; TPT) transmitted by the 925VST when it is in the SMOS Responder
mode. TPT is transmitted by the SMOS Responder function in response to a “call” from
the far end SMOS Director function. The duration can be from 1 to 99 seconds; the default
is 10 seconds.
N
O
A
NSWER
T
IMEOUT
While the 925VST dials the responder to begin a test, the display reads
Calling Test
Line, Please wait…
You can specify how long the 925VST will wait for an
outgoing call to be answered. If the responder does not answer the call, the test is aborted
and the display reads
Test Aborted
. At the No Answer Timeout screen, enter the
number of seconds the 925VST waits for an outgoing call to be answered. The maximum
wait allowed is 99 seconds; the default is 40 seconds.
M
AX
W
AIT
D
IAL
-T
ONE
When the 925VST is connected to the line under test and the desired responder phone
number is selected, the 925VST listens for a dial tone before dialing the responder. The
925VST dials the responder number even if no dial tone is detected. At the Max Wait
Dial-Tone screen, enter the desired number of seconds the 925VST will wait for a dial
tone before dialing the responder. The maximum wait allowed is 99 seconds; the default is
10 seconds.
A
UDIBLE
A
LERT
Audible alerts indicate test activities such as an aborted test or test has failed. At the
Audible Alert screen, press 1 to enable or 2 to disable the audible alerts. The default is
Disabled.
B
ACKLIGHT
T
IMEOUT
The backlight illuminates the display when you press the light bulb key. At the Backlight
Timeout screen, enter the number of seconds the display light is on before it automatically
shuts off. The maximum illumination time is 99 seconds; the default is 15 seconds.