3-5
Hazardous Area Oxymitter 5000
For optimum performance, Emerson Process Management recommends
connecting the output to a Potter & Brumfield 3.2 mA DC relay
(P/N R10S-E1Y1-J1.0K).
Of the ten modes in Table 3-1, modes 0 through 7 are the alarm modes. The
factory default is mode 5 for Hazardous Area Oxymitter 5000 units without an
IMPS 4000 or SPS 4001B. In this mode, the output will signal when a unit
alarm or a CALIBRATION RECOMMENDED indication occurs.
Calibration Handshake Signal
If using an optional IMPS 4000 or SPS 4001B, the logic I/O must be
configured for calibration handshaking. Of the ten modes in Table 3-1 only
modes 8 and 9 are configured for calibration handshaking. For a Hazardous
Area Oxymitter 5000 with an IMPS 4000 or an SPS 4001B, the factory sets
the default to mode 8. In this mode, the logic I/O will be used to communicate
between the Hazardous Area Oxymitter 5000 and the sequencer and to signal
the sequencer when a CALIBRATION RECOMMENDED indication occurs.
Table 3-1. Logic I/O
Configuration
*The default condition for a Hazardous Area Oxymitter 5000 without an IMPS 4000 or
SPS 4001B.
**The default condition for a Hazardous Area Oxymitter 5000 with an IMPS 4000 or SPS 4001B.
Table 3-2. Logic I/O Parameters
Mode
Configuration
0
The unit is not configured for any alarm condition.
1
The unit is configured for a Unit Alarm.
2
The unit is configured for Low O
2
.
3
The unit is configured for both a Unit Alarm and Low O
2
.
4
The unit is configured for a High AC Impedance/CALIBRATION
RECOMMENDED.
5*
The unit is configured for both a Unit Alarm and a High AC
Impedance/CALIBRATION RECOMMENDED.
6
The unit is configured for both a Low O
2
and High AC Impedance/CALIBRATION
RECOMMENDED.
7
The unit is configured for a Unit Alarm, a Low O
2
, and a High AC
Impedance/CALIBRATION RECOMMENDED.
8**
The unit is configured for a calibration handshake with IMPS 4000 or SPS 4001B.
CALIBRATION RECOMMENDED will initiate the calibration cycle.
9
The unit is configured for a calibration handshake. CALIBRATION
RECOMMENDED will not initiate the calibration cycle with the IMPS 4000 or SPS
4001B.
Parameter
Definition
Range
Parameter
Number
IO_PIN_MODE
This parameter represents the operating
mode of the discrete IO pin of the
transmitter.
1-10
40
IO_PIN_STATE
This parameter represents the current state
of the transmitter’s discrete IO pin.
0=FALSE, 1=TRUE.
0-1
41
Содержание Oxymitter 5000
Страница 2: ......
Страница 6: ......
Страница 14: ...Hazardous Area Oxymitter 5000 iv Instruction Manual IM 106 350C Rev 2 2 July 2008 ...
Страница 82: ...Hazardous Area Oxymitter 5000 7 6 Instruction Manual IM 106 350C Rev 2 2 July 2008 ...
Страница 130: ...Hazardous Area Oxymitter 5000 9 24 Instruction Manual IM 106 350C Rev 2 2 July 2008 ...
Страница 140: ...Hazardous Area Oxymitter 5000 11 6 Instruction Manual IM 106 350C Rev 2 2 July 2008 ...
Страница 172: ...Hazardous Area Oxymitter 5000 B 2 Instruction Manual IM 106 350C Rev 2 2 July 2008 ...
Страница 192: ...Hazardous Area Oxymitter 5000 D 14 Instruction Manual IM 106 350C Rev 2 2 July 2008 ...
Страница 210: ...Hazardous Area Oxymitter 5000 E 18 Instruction Manual IM 106 350C Rev 2 2 July 2008 ...
Страница 212: ...Instruction Manual IM 106 350C Rev 2 2 July 2008 Hazardous Area Oxymitter 5000 Index 2 ...