Hazardous Area Oxymitter 5000
BLOCK_ALM
44
None
The block alarm is used for all configuration, hardware, connection failure, or
system problems in the block. The cause of the alert is entered in the
subcode field. The first alert to become active will set the active status in the
status parameter. As soon as the Unreported status is cleared by the alert
reporting task, and other block alert may be reported without clearing the
Active status, if the subcode has changed.
BLOCK_ERR
06
None
This parameter reflects the error status associated with the hardware or
software components associated with a block. It is a bit string so that multiple
errors may be shown.
BYPASS
17
None
Used to override the calculation of the block. When enabled, the SP is sent
directly to the output.
CAS_IN
18
EU of PV_SCALE
The remote setpoint value from another block.
CONTROL_OPTS
13
None
Allows you to specify control strategy options. The supported control options
for the PID block are Track enable, Track in Manual, SP-PV Track in Man,
SP-PV Track in LO or IMAN, Use PV for BKCAL OUT, and Direct Acting.
CONTROL_OPTS
13
None
Allows you to specify control strategy options. The supported control options
for the PID block are Track enable, Track in Manual, SP-PV Track in Man,
SP-PV Track in LO or IMAN, Use PV for BKCAL OUT, and Direct Acting.
DV_HI_ALM
64
None
The DV HI alarm data, which includes a value of the alarm, a timestamp of
occurrence, and the state of the alarm.
DV_HI_LIM
57
EU of PV_SCALE
The setting for the alarm limit used to detect the deviation high alarm
condition.
DV_HI_PRI
56
None
The priority of the deviation high alarm.
DV_LO_ALM
65
None
The DV LO alarm data, which includes a value of the alarm, a timestamp of
occurrence, and the state of the alarm.
DV_LO_LIM
59
EU of PV_SCALE
The setting for the alarm limit use to detect the deviation low alarm condition.
DV_LO_PRI
58
None
The priority of the deviation low alarm.
ERROR
67
EU of PV_SCALE
The error (SP-PV) used to determine the control action.
FF_ENABLE
70
None
Enables the use of feedforward calculations.
FF_GAIN
42
None
The feedforward gain value. FF_VAL is multiplied by FF_GAIN before it is
added to the calculated control output.
FF_SCALE
41
None
The high and low scale values, engineering units code, and number of digits
to the right of the decimal point associated with the feedforward value
(FF_VAL).
FF_VAL
40
EU of FF_SCALE
The feedforward control input value and status.
GAIN
23
None
The proportional gain value. This value cannot = 0.
GRANT_DENY
12
None
Options for controlling access of host computers and local control panels to
operating, tuning, and alarm parameters of the block. Not used by the device.
HI_ALM
61
None
The HI alarm data, which includes a value of the alarm, a timestamp of
occurrence, and the state of the alarm.
HI_HI_ALM
60
None
The HI HI alarm data, which includes a value of the alarm, a timestamp of
occurrence, and the state of the alarm.
HI_HI-LIM
49
EU of PV_SCALE
The setting for the alarm limit used to detect the HI HI alarm condition.
HI_HI_PRI
48
None
The priority of the HI HI Alarm.
HI_LIM
51
EU of PV_SCALE
The setting for the alarm limit used to detect the HI alarm condition.
HI_PRI
50
None
The priority of the HI alarm.
IN
15
EU of PV_SCALE
The connection for the PV input from another block.
LO_ALM
62
None
The LO alarm data, which includes a value of the alarm, a timestamp of
occurrence, and the state of the alarm.
LO_LIM
53
EU of PV_SCALE
The setting for the alarm limit used to detect the LO alarm condition.
LO_LO_ALM
63
None
The LO LO alarm data, which includes a value of the alarm, a timestamp of
occurrence, and the state of the alarm.
LO_LO_LIM
55
EU of PV_SCALE
The setting for the alarm limit used to detect the LO LO alarm condition.
Parameter
Index
Number
Units
Description
Содержание Oxymitter 5000
Страница 2: ......
Страница 6: ......
Страница 14: ...Hazardous Area Oxymitter 5000 iv Instruction Manual IM 106 350C Rev 2 2 July 2008 ...
Страница 82: ...Hazardous Area Oxymitter 5000 7 6 Instruction Manual IM 106 350C Rev 2 2 July 2008 ...
Страница 130: ...Hazardous Area Oxymitter 5000 9 24 Instruction Manual IM 106 350C Rev 2 2 July 2008 ...
Страница 140: ...Hazardous Area Oxymitter 5000 11 6 Instruction Manual IM 106 350C Rev 2 2 July 2008 ...
Страница 172: ...Hazardous Area Oxymitter 5000 B 2 Instruction Manual IM 106 350C Rev 2 2 July 2008 ...
Страница 192: ...Hazardous Area Oxymitter 5000 D 14 Instruction Manual IM 106 350C Rev 2 2 July 2008 ...
Страница 210: ...Hazardous Area Oxymitter 5000 E 18 Instruction Manual IM 106 350C Rev 2 2 July 2008 ...
Страница 212: ...Instruction Manual IM 106 350C Rev 2 2 July 2008 Hazardous Area Oxymitter 5000 Index 2 ...