![Rosemount Oxymitter 5000 Скачать руководство пользователя страница 40](http://html1.mh-extra.com/html/rosemount/oxymitter-5000/oxymitter-5000_instruction-manual_1487464040.webp)
Hazardous Area Oxymitter 5000
2-8
Instruction Manual
IM-106-350C, Rev 2.2
July 2008
4. If using the optional ceramic diffusion element, the vee deflector must
be correctly oriented. Before inserting the Hazardous Area Oxymitter
5000, check the direction of flow of the gas in the duct. Orient the vee
deflector so the apex points upstream toward the flow (Figure 2-6). This
may be done by loosening the setscrews and rotating the vee deflector
to the desired position. Retighten the setscrews.
5. In vertical installations, ensure the system cable drops vertically from
the Hazardous Area Oxymitter 5000 and the conduit is routed below the
level of the electronics housing. This drip loop minimizes the possibility
that moisture will damage the electronics (Figure 2-7).
6. If the system has an abrasive shield, check the dust seal gaskets. The
joints in the two gaskets must be staggered 180 degrees. Make sure the
gaskets are in the hub grooves as the Hazardous Area Oxymitter 5000
slides into the 15 degree forcing cone in the abrasive shield.
NOTE
If process temperatures will exceed 200°C (392°F), use anti-seize compound
on the stud threads to ease future removal of the Hazardous Area Oxymitter
5000.
7. Insert probe through the opening in the mounting plate and bolt the unit
to the plate.
NOTE
To maintain CE compliance, ensure a good connection exists between the
mounting plate studs or earthing screws on electronics housing and earth.
8. Ensure the Hazardous Area Oxymitter 5000 is properly earthed by way
of both internal and external points.
9. If duct work insulation is removed for Hazardous Area Oxymitter 5000
probe mounting, make sure the insulation is replaced afterward. See
Figure 2-7.
NOTE
For probe temperatures that will exceed 85°C (185°F ), we recommend the
remote mounted electronics option.
10. Ensure the probe installation does not obscure the warnings on the
housing covers.
Uninsulated stacks or ducts may cause ambient temperatures around the electronics to
exceed 85°C (185°F), which may cause overheating da mage to the electronics
.
Содержание Oxymitter 5000
Страница 2: ......
Страница 6: ......
Страница 14: ...Hazardous Area Oxymitter 5000 iv Instruction Manual IM 106 350C Rev 2 2 July 2008 ...
Страница 82: ...Hazardous Area Oxymitter 5000 7 6 Instruction Manual IM 106 350C Rev 2 2 July 2008 ...
Страница 130: ...Hazardous Area Oxymitter 5000 9 24 Instruction Manual IM 106 350C Rev 2 2 July 2008 ...
Страница 140: ...Hazardous Area Oxymitter 5000 11 6 Instruction Manual IM 106 350C Rev 2 2 July 2008 ...
Страница 172: ...Hazardous Area Oxymitter 5000 B 2 Instruction Manual IM 106 350C Rev 2 2 July 2008 ...
Страница 192: ...Hazardous Area Oxymitter 5000 D 14 Instruction Manual IM 106 350C Rev 2 2 July 2008 ...
Страница 210: ...Hazardous Area Oxymitter 5000 E 18 Instruction Manual IM 106 350C Rev 2 2 July 2008 ...
Страница 212: ...Instruction Manual IM 106 350C Rev 2 2 July 2008 Hazardous Area Oxymitter 5000 Index 2 ...