Hazardous Area Oxymitter 5000
9-8
Instruction Manual
IM-106-350C, Rev 2.2
July 2008
HAZARDOUS AREA
OXYMITTER 5000
REPAIR
Each of the following procedures details how to remove and replace a specific
component of the Hazardous Area Oxymitter 5000.
Removal and
Replacement of Probe
1. Remove.
a. Turn off power to the system.
b. Shut off the calibration gases at the cylinders and the instrument air.
c. Disconnect the calibration gas and instrument air lines from the
Hazardous Area Oxymitter 5000.
d. While facing the Hazardous Area Oxymitter 5000 and looking at the
Rosemount label, remove screw (18, Figure 9-3 or Figure 9-4), cover
lock (19), and captive washer (20) securing cover (17) on left side of
housing (11). Remove the cover to expose terminal block (15).
e. Remove all signal and power wiring to the probe.
f. Remove insulation to access the mounting bolts.
g. Unbolt the Hazardous Area Oxymitter 5000 from the stack and take it
to a clean work area.
h. Allow the unit to cool to a comfortable working temperature.
2. Replace.
a. Bolt the Hazardous Area Oxymitter 5000 to the stack and install the
insulation.
b. Connect all signal and power leads at the probe. Refer to Section 2:
Installation, for detailed wiring instructions.
c. Install left side cover (17, Figure 9-3 or Figure 9-4) and ensure it is
tight. Secure the cover using captive washer (20), cover lock (19),
and screw (18).
d. Connect the calibration gas and instrument air lines to probe.
e. Turn on the calibration gases at the cylinders and turn on instrument
air.
f. Restore power to the system; refer to "Power Up" in Section 5:
Startup and Operation of Hazardous Area Oxymitter 5000 with
Membrane Keypad or "Power Up" in Section 6: Startup and
Operation of Hazardous Area Oxymitter 5000 with LOI. When the
probe is at operating temperature, calibrate the probe per
"Calibration with Keypad".
It is recommended that the Hazardous Area Oxymitter 5000 be removed from the stack for
all service activities. The unit should be allowed to cool and be taken to a clean work area.
Failure to comply may cause severe burns.
Disconnect and lock out power before working on any electrical components. There is
voltage up to 115 VAC.
Содержание Oxymitter 5000
Страница 2: ......
Страница 6: ......
Страница 14: ...Hazardous Area Oxymitter 5000 iv Instruction Manual IM 106 350C Rev 2 2 July 2008 ...
Страница 82: ...Hazardous Area Oxymitter 5000 7 6 Instruction Manual IM 106 350C Rev 2 2 July 2008 ...
Страница 130: ...Hazardous Area Oxymitter 5000 9 24 Instruction Manual IM 106 350C Rev 2 2 July 2008 ...
Страница 140: ...Hazardous Area Oxymitter 5000 11 6 Instruction Manual IM 106 350C Rev 2 2 July 2008 ...
Страница 172: ...Hazardous Area Oxymitter 5000 B 2 Instruction Manual IM 106 350C Rev 2 2 July 2008 ...
Страница 192: ...Hazardous Area Oxymitter 5000 D 14 Instruction Manual IM 106 350C Rev 2 2 July 2008 ...
Страница 210: ...Hazardous Area Oxymitter 5000 E 18 Instruction Manual IM 106 350C Rev 2 2 July 2008 ...
Страница 212: ...Instruction Manual IM 106 350C Rev 2 2 July 2008 Hazardous Area Oxymitter 5000 Index 2 ...