Instruction Manual
IM-106-350C, Rev 2.2
July 2008
Hazardous Area Oxymitter 5000
http://www.raihome.com
Appendix C
Fieldbus Parameter Description
FIELDBUS PARAMETERS
Parameter Mnemonic
Valid Range
Initial
Value
Units
Description
Parameter
Number
ALARM_POINT_LOW
0.0-40.0
% O2
This is the point at which the Low O
2
alarm
will become active.
37
ALERT_KEY
See Function Block Specification, part 1.
FF-890, page 50.
4
AUTOCAL_ENABLED
0: Not enabled
1: Enabled
Enumerated
This parameter enables automatic
calibrations.
46
AUTOCAL_INTERVAL
0-9999
Hours
This is the time between automatic
calibrations, in hours.
47
BLOCK_ALM
See Function Block Specification, part 2.
FF-891, page 41.
8
BLOCK_ERR
See Function Block Specification, part 2.
FF-891, page 41.
6
BUILD_DATE
0
N/A
This is the date that the Oxymitter software
was built.
62
BUILD_NUMBER
0-65535
0
N/A
This is the build number of the Oxymitter
software.
60
CAL_CONSTANT
± 20.0
mV
This parameter represents the constant (off-
set) value used in the calculation of convert-
ing the sensor voltage to an O
2
value. The
value of this parameter may be manually
entered or calculated during a sensor cali-
bration.
27
CAL_GAS_TIME
60-1200
Sec
This is the length of time test gases are
applied to the O2 probe before each
measurement of the calibration cycle.
22
CAL_LAST_
CONSTANT
± 20.0
mV
The constant from the last successful
calibration.
29
CAL_LAST_SLOPE
34.5-57.5
mV/Decade
The slope from the last successful
calibration.
28
CAL_POINT_HI
0.0-40.0
% O2
This is the actual value of the gas being
applied during the high test gas phase of a
calibration.
17
CAL_POINT_LO
0.0-40.0
% O2
This is the actual value of the gas being
applied during the low test gas phase of a
calibration.
18
CAL_PURGE_TIME
60-1200
Sec
This is the length of time after a calibration is
complete before the O2 value status returns
to normal.
23
CAL_SENSOR_
IMPEDANCE
0-10000
Ohms
This is the sensor impedance value
measured at the time of the last calibration.
25
Содержание Oxymitter 5000
Страница 2: ......
Страница 6: ......
Страница 14: ...Hazardous Area Oxymitter 5000 iv Instruction Manual IM 106 350C Rev 2 2 July 2008 ...
Страница 82: ...Hazardous Area Oxymitter 5000 7 6 Instruction Manual IM 106 350C Rev 2 2 July 2008 ...
Страница 130: ...Hazardous Area Oxymitter 5000 9 24 Instruction Manual IM 106 350C Rev 2 2 July 2008 ...
Страница 140: ...Hazardous Area Oxymitter 5000 11 6 Instruction Manual IM 106 350C Rev 2 2 July 2008 ...
Страница 172: ...Hazardous Area Oxymitter 5000 B 2 Instruction Manual IM 106 350C Rev 2 2 July 2008 ...
Страница 192: ...Hazardous Area Oxymitter 5000 D 14 Instruction Manual IM 106 350C Rev 2 2 July 2008 ...
Страница 210: ...Hazardous Area Oxymitter 5000 E 18 Instruction Manual IM 106 350C Rev 2 2 July 2008 ...
Страница 212: ...Instruction Manual IM 106 350C Rev 2 2 July 2008 Hazardous Area Oxymitter 5000 Index 2 ...