![Rosemount Oxymitter 5000 Скачать руководство пользователя страница 121](http://html1.mh-extra.com/html/rosemount/oxymitter-5000/oxymitter-5000_instruction-manual_1487464121.webp)
Instruction Manual
IM-106-350C, Rev 2.2
July 2008
9-15
Hazardous Area Oxymitter 5000
Figure 9-7. Fuse Location
10. Reconnect the J8 connector to the power supply board. Make sure the
connector is secure.
11. Holding the J1 connector leads, slide the electronic assembly the rest of
the way into the housing. Seat the electronic assembly on the mating
connector pins.
12. Gently try to rotate the electronic assembly to check for full seating. If
the electronic assembly rotates, remove the assembly and repeat step
11.
13. Reconnect the J1 connector to the microprocessor board (Figure 9-5).
Ensure the connector is secure.
14. Tighten three captive screws (10, Figure 9-3 or Figure 9-4) in the top of
microprocessor board (5).
15. Replace right housing cover (1, Figure 9-3); make sure it is tight. Secure
the cover using captive washer (20), cover lock (19), and screw (18).
16. Install and tighten cover (1); make sure it is tight. Secure the cover using
captive washer (20), cover lock (19), and screw (18).
Entire Probe
Replacement (Excluding
Electronics)
Do not attempt to replace the probe until all other possibilities for poor
performance have been considered. If probe replacement is needed, see
Table 10-1 for part numbers.
Fuse
Power
Supply
Board
38740020
When working on this equipment on the laboratory bench, be aware that the Hazardous
Area Oxymitter 5000, probe tube, and flame arrestor hub can be hot [up to 300°C (572°F)]
in the region of the probe heater.
Содержание Oxymitter 5000
Страница 2: ......
Страница 6: ......
Страница 14: ...Hazardous Area Oxymitter 5000 iv Instruction Manual IM 106 350C Rev 2 2 July 2008 ...
Страница 82: ...Hazardous Area Oxymitter 5000 7 6 Instruction Manual IM 106 350C Rev 2 2 July 2008 ...
Страница 130: ...Hazardous Area Oxymitter 5000 9 24 Instruction Manual IM 106 350C Rev 2 2 July 2008 ...
Страница 140: ...Hazardous Area Oxymitter 5000 11 6 Instruction Manual IM 106 350C Rev 2 2 July 2008 ...
Страница 172: ...Hazardous Area Oxymitter 5000 B 2 Instruction Manual IM 106 350C Rev 2 2 July 2008 ...
Страница 192: ...Hazardous Area Oxymitter 5000 D 14 Instruction Manual IM 106 350C Rev 2 2 July 2008 ...
Страница 210: ...Hazardous Area Oxymitter 5000 E 18 Instruction Manual IM 106 350C Rev 2 2 July 2008 ...
Страница 212: ...Instruction Manual IM 106 350C Rev 2 2 July 2008 Hazardous Area Oxymitter 5000 Index 2 ...