Hazardous Area Oxymitter 5000
9-6
Instruction Manual
IM-106-350C, Rev 2.2
July 2008
FOUNDATION FIELDBUS
O
2
CAL METHOD
To perform a calibration using the FOUNDATION fieldbus use the following
procedure.
1. From the computer running the fieldbus control program run the O
2
Cal
Method.
2. In the first O
2
CAL screen, a "Loop should be removed from automatic
control" warning appears. Remove the Oxymitter 5000 from any
automatic control loops to avoid a potentially dangerous operating
condition and press OK.
3. From this point, following the on-screen prompts to complete the
calibration procedure. When a step is complete, select Proceed to Next
Step and press the Next button.
4. During the wait periods, such as during a purge, the Time Remaining
display may be updated by selecting Update and pressing Next button.
CALIBRATION WITH LOI
Refer to Figure 6-4 for a view of the LOI menu tree. To calibrate the
Hazardous Area Oxymitter 5000 from the LOI, access the CALIBRATION/
Start Calibration menu.
CALIBRATION/Start Calibration
This is the starting point for calibrations. The LOI will instruct the user through
this entire procedure. You can select "Abort Calib" at any time to abort the
calibration.
1. The LOI displays the following:
Apply Gas 1
Hit E when ready
The Oxymitter 5000 is ready to accept the first calibration gas. Apply the
first calibration gas. (Electronics will abort the calibration if this step is
not done within 30 minutes).
2. Touch the Enter key to start the Gas 1 flow. A timer is activated to allow
the calibration gas adequate time to flow (default time of five minutes).
The LOI displays:
Flow Gas 1 xxxxs
Read Gas 1 xxxxs
Done Gas 1
The display counts down the seconds remaining to flow Gas 1 then the
time remaining for sensing the O
2
concentration of Gas 1. Done Gas 1
indicates completion.
Failure to remove the Oxymitter 5000 from automatic control loops prior to performing this
procedure may result in dangerous operating conditions.
Содержание Oxymitter 5000
Страница 2: ......
Страница 6: ......
Страница 14: ...Hazardous Area Oxymitter 5000 iv Instruction Manual IM 106 350C Rev 2 2 July 2008 ...
Страница 82: ...Hazardous Area Oxymitter 5000 7 6 Instruction Manual IM 106 350C Rev 2 2 July 2008 ...
Страница 130: ...Hazardous Area Oxymitter 5000 9 24 Instruction Manual IM 106 350C Rev 2 2 July 2008 ...
Страница 140: ...Hazardous Area Oxymitter 5000 11 6 Instruction Manual IM 106 350C Rev 2 2 July 2008 ...
Страница 172: ...Hazardous Area Oxymitter 5000 B 2 Instruction Manual IM 106 350C Rev 2 2 July 2008 ...
Страница 192: ...Hazardous Area Oxymitter 5000 D 14 Instruction Manual IM 106 350C Rev 2 2 July 2008 ...
Страница 210: ...Hazardous Area Oxymitter 5000 E 18 Instruction Manual IM 106 350C Rev 2 2 July 2008 ...
Страница 212: ...Instruction Manual IM 106 350C Rev 2 2 July 2008 Hazardous Area Oxymitter 5000 Index 2 ...