Recalibration is required whenever electronic cards or sensing cell is
replaced.
Replacement of Entire
Electronics (with
Housing)
1. Follow the instructions in "Removal and Replacement of Probe" to
remove the Hazardous Area Oxymitter 5000 probe from the stack or
duct.
2. Remove four screws (22, Figure 9-3) and washers (21) from probe tube
assembly (23). Remove the probe tube assembly from the housing (11).
3. Disconnect the heater and signal wire connectors from the mating
connectors on the heater strut assembly (36).
NOTE
The integral electrical barrier/feed-through is thread-locked into the electrical
housing and cannot be removed.
4. Make sure the o-ring (35) is in good condition. Replace o-ring if
damaged.
5. Make sure that the conduit port of the electronic housing is on the same
side as the CAL and REF gas ports. Install four washers (21) and
screws (22). Tighten screws.
6. Follow the instructions in "Probe Installation" in Section 2: Installation to
install the Hazardous Area Oxymitter 5000 into the stack or duct.
Do not force the probe housing when installing or removing from the integral electrical
barrier/feed-through (Figure 9-3). Damage to the aluminum probe housing can occur.
Opening the electronic housing will cause the loss of ALL hazardous permits. Opening the
electronics housing in hazardous areas may cause an explosion resulting in loss of
property, severe personal injury, or death. It may be required to get a hot work permit from
your company safety officer before opening the electronic housing.
Содержание Oxymitter 5000
Страница 2: ......
Страница 6: ......
Страница 14: ...Hazardous Area Oxymitter 5000 iv Instruction Manual IM 106 350C Rev 2 2 July 2008 ...
Страница 82: ...Hazardous Area Oxymitter 5000 7 6 Instruction Manual IM 106 350C Rev 2 2 July 2008 ...
Страница 130: ...Hazardous Area Oxymitter 5000 9 24 Instruction Manual IM 106 350C Rev 2 2 July 2008 ...
Страница 140: ...Hazardous Area Oxymitter 5000 11 6 Instruction Manual IM 106 350C Rev 2 2 July 2008 ...
Страница 172: ...Hazardous Area Oxymitter 5000 B 2 Instruction Manual IM 106 350C Rev 2 2 July 2008 ...
Страница 192: ...Hazardous Area Oxymitter 5000 D 14 Instruction Manual IM 106 350C Rev 2 2 July 2008 ...
Страница 210: ...Hazardous Area Oxymitter 5000 E 18 Instruction Manual IM 106 350C Rev 2 2 July 2008 ...
Страница 212: ...Instruction Manual IM 106 350C Rev 2 2 July 2008 Hazardous Area Oxymitter 5000 Index 2 ...