Hazardous Area Oxymitter 5000
Situation #2
The transmitter in situation #1 is installed below the tank in a position where
the liquid column in the impulse line, when the tank is empty, is equivalent to
2.0 psi (Figure D-6).
Figure D-6. Situation #2
Diagram
Solution
Table D-5 lists the appropriate configuration settings.
Table D-5. Analog Input
Function Diagram for a Pressure
Transmitter used in Level
Measurement (Situation #2)
38740088
16 ft
0 ft
2.0 psi measured at
the transmitter
Empty Tank
Parameter
Configured Values
L_TYPE
Indirect
XD_SCALE
2 to 9 psi
OUT_SCALE
0 to 16 ft
Содержание Oxymitter 5000
Страница 2: ......
Страница 6: ......
Страница 14: ...Hazardous Area Oxymitter 5000 iv Instruction Manual IM 106 350C Rev 2 2 July 2008 ...
Страница 82: ...Hazardous Area Oxymitter 5000 7 6 Instruction Manual IM 106 350C Rev 2 2 July 2008 ...
Страница 130: ...Hazardous Area Oxymitter 5000 9 24 Instruction Manual IM 106 350C Rev 2 2 July 2008 ...
Страница 140: ...Hazardous Area Oxymitter 5000 11 6 Instruction Manual IM 106 350C Rev 2 2 July 2008 ...
Страница 172: ...Hazardous Area Oxymitter 5000 B 2 Instruction Manual IM 106 350C Rev 2 2 July 2008 ...
Страница 192: ...Hazardous Area Oxymitter 5000 D 14 Instruction Manual IM 106 350C Rev 2 2 July 2008 ...
Страница 210: ...Hazardous Area Oxymitter 5000 E 18 Instruction Manual IM 106 350C Rev 2 2 July 2008 ...
Страница 212: ...Instruction Manual IM 106 350C Rev 2 2 July 2008 Hazardous Area Oxymitter 5000 Index 2 ...