Reference Manual
00809-0100-4848, Rev AA
July 2009
Rosemount 848T Wireless
www.rosemount.com
Appendix A
Specifications and
Reference Data
Specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . page A-1
Dimensional Drawings . . . . . . . . . . . . . . . . . . . . . . . . . . . . page A-5
Ordering Information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . page A-6
SPECIFICATIONS
Functional
Specifications
Input
Supports Thermocouple, RTD, millivolt, ohm, and milliamp input types. See
Accuracy on page A-3 for sensor options.
Output
WirelessHART
™
, linear with temperature or input.
Humidity Limits
0 – 99% non-condensing relative humidity.
Transmit Rate
User selectable, 8 sec. to 60 min.
Accuracy
(PT 100 @ reference conditions: 20 °C)
±0.30 °C (±0.54 °F)
Physical Specifications
Electrical Connections/Power Module
• Replaceable, non-rechargeable, Intrinsically Safe Lithium-Thionyl
Chloride power module with polybutadine terephthalate (PBT)
enclosure.
• Six year life at reference conditions.
(1)
• 5-Screw Terminals for sensor connection.
Field Communicator Connections
Clips permanently fixed to terminal block
Materials of Construction
Housing
• Low-copper aluminum
Paint
• Polyurethane
(1)
Reference conditions are 70° F (21° C), transmit rate of once per minute, and routing data
for three additional network devices.
Содержание 848T
Страница 2: ......
Страница 12: ...Reference Manual 00809 0100 4848 Rev AA July 2009 Rosemount 848T Wireless 1 6 ...
Страница 32: ...Reference Manual 00809 0100 4848 Rev AA July 2009 Rosemount 848T Wireless 3 12 ...
Страница 40: ...Reference Manual 00809 0100 4848 Rev AA July 2009 Rosemount 848T Wireless 5 4 ...
Страница 44: ...Reference Manual 00809 0100 4848 Rev AA July 2009 Rosemount 848T Wireless 6 4 ...
Страница 59: ......