DEFAULT SETTINGS
The 848T default configuration is shown below:
Use the C1 option code to have the factory configure each sensor individually.
This option also enables factory configuration of process alerts, update rate,
and channel tag. This option code is not required to configure the
self-organizing network parameters, or to set all of the sensors identically.
DEVICE NETWORK
CONFIGURATION
Join Device to Network
The transmitter must be configured in order to communicate with the Smart
Wireless Gateway, and ultimately with the host system. This step is the
wireless equivalent of connecting wires from the transmitter to the host
system.
1. From the
Home
screen, select 2:
Configure.
2. Select 1:
Guided Setup
.
3. Select 1:
Join Device to Network
.
Using a Field Communicator or AMS, enter the Network ID and Join Key so
they match the Network ID and Join Key of the Smart Wireless Gateway, and
other devices in the network. If the Network ID and Join Key do not match the
Gateway, the transmitter will not communicate with the network. The Network
ID and Join Key can be obtained from the Smart Wireless Gateway on the
Setup>Network>Settings
page on the web server.
Sensor 1
Type J Thermocouple
Sensor 2
Type J Thermocouple
Sensor 3
Type J Thermocouple
Sensor 4
Type J Thermocouple
Engineering Units
°C
Number of Lead Wires
2
Sensor Alerts
Disabled
Network ID
Factory Generated Network Parameters
Join Key
Factory Generated Network Parameters
Update Rate
1 Minute
Fast Key
2, 1, 1
Содержание 848T
Страница 2: ......
Страница 12: ...Reference Manual 00809 0100 4848 Rev AA July 2009 Rosemount 848T Wireless 1 6 ...
Страница 32: ...Reference Manual 00809 0100 4848 Rev AA July 2009 Rosemount 848T Wireless 3 12 ...
Страница 40: ...Reference Manual 00809 0100 4848 Rev AA July 2009 Rosemount 848T Wireless 5 4 ...
Страница 44: ...Reference Manual 00809 0100 4848 Rev AA July 2009 Rosemount 848T Wireless 6 4 ...
Страница 59: ......