R&S®SML / R&S®SMV03
Test sequence
1090.3123.12 10.7
E-6
- FREQ/CENTER/STOP FREQUENCY,
- FM DEMOD ON,
- Demodulation bandwidth 100 kHz,
- Span 200 Hz,
- MEAS TIME 10 ms.
Settings on R&S SML /
R&S SMV03
- LEVEL: 0 dBm
- UTILITIES REF OSC SOURCE EXTERN
Measurement
Setting on spectrum analyzer: TRIGGER EXTERN
- external triggering on positive edge at 1.4 V
Send target frequency from process controller.
The build-up trace appears on the screen of the externally
triggered spectrum analyzer. The settling time can be determined
with the aid of the cursor.
Repeat measurement with start and target frequencies interchanged.
The following hops need to be measured:
Table 10-3 Measuring the frequency setting time
Start frequency
Target frequency
806 MHz
808 MHz
606 MHz
1075 MHz
1075 MHz
76 MHz
76 MHz
1075 MHz
1210.5 MHz
2200.000001 MHz
2800 MHz
1818 MHz