R&S®SML / R&S®SMV03
Test sequence
1147.7621.12 10.47
E-6
I/Q Imbalance
Measurement of Imbalance
Test equipment
Spectrum analyzer (Table 5-1, item 2)
Adjustable DC voltage source (Table 5-1, item 12).
Test setup
Connect the spectrum analyzer to the RF output of the R&S SMV03.
Connect the DC voltage source to the I or the Q input.
Measurement
Settings on the R&S SMV03
Test frequencies: 250MHz, 0.1Hz, 420MHz,
0.1Hz, 665MHz, 900 MHz, 1170MHz, 1900MHz,
3000MHz
Test level 0 dBm
Select State IQ in the menu vector Mod.
Settings on analyzer
Center frequency = test frequency., Span 1 MHz
Reference level = test level +3 dB
Scale 1 dB/div.
First measure the undistorted level as a reference. To this end,
apply a DC voltage of 0.500 V to the I and then to the Q and note
down the corresponding RF levels as reference levels. In the menu
Vector Mod/Impairment State select On and Imbalance 10 %.
repeat the Level measurement.
The I level should increase by the set imbalance, the Q level
decrease by the inverse ratio. With 10 %, the I level should
increase to 1.1 times, the Q level should be reduced to 0.909 times
the original value (corresponding to
±
0.83 dB).
Содержание SML01
Страница 4: ......
Страница 14: ......
Страница 22: ......
Страница 32: ......
Страница 34: ......
Страница 36: ......
Страница 38: ......
Страница 40: ......
Страница 42: ......
Страница 44: ......
Страница 56: ......
Страница 62: ......
Страница 77: ...R S SML R S SMV03 Menu Summary for R S SML 1090 3123 12 E 6 3 15 Menu Summary for R S SML ...
Страница 78: ...Menu Summary for R S SMV03 R S SML R S SMV03 1090 3123 12 E 6 3 16 Menu Summary for R S SMV03 ...
Страница 148: ......
Страница 240: ......
Страница 242: ......