Sweep
R&S®SML / R&S®SMV03
1090.3123.12
E-6
4.54
LF Sweep
Settings for LF sweeps can be made in the Sweep - LFGen menu.
Menu selection: Sweep - LFGen
Fig. 4-21
Sweep - LFGen menu
Start Freq
Input value of start frequency.
IEC/IEEE-bus command
:SOUR2:FREQ:STAR 1kHz
Stop Freq
Input value of stop frequency.
IEC/IEEE-bus command
:SOUR2:FREQ:STOP 100kHz
Current Freq
Display of current frequency value.
In Step mode: input value of frequency.
Spacing
Selection of linear or logarithmic sweep.
IEC/IEEE-bus command
:SOUR2:SWE:SPAC LIN
Step Lin
Input value of step size.
IEC/IEEE-bus command
:SOUR2:SWE:STEP:LIN 1kHz
Dwell
Input value of dwell time per step.
IEC/IEEE-bus command
:SOUR2:SWE:DWEL 15ms
Mode
Selection of sweep mode (see "Operating Modes").
IEC/IEEE-bus command
:SOUR2:FREQ:MODE SWE;
:SOUR2:SWE:MODE AUTO;
:TRIG2:SOUR SING
Reset Sweep
Sets the start frequency.
IEC/IEEE-bus command
:ABOR
Exec Single Sweep
Starts a single sweep. This function is displayed and is effective only if Single
Mode is selected.
IEC/IEEE-bus command
:TRIG
Содержание SML01
Страница 4: ......
Страница 14: ......
Страница 22: ......
Страница 32: ......
Страница 34: ......
Страница 36: ......
Страница 38: ......
Страница 40: ......
Страница 42: ......
Страница 44: ......
Страница 56: ......
Страница 62: ......
Страница 77: ...R S SML R S SMV03 Menu Summary for R S SML 1090 3123 12 E 6 3 15 Menu Summary for R S SML ...
Страница 78: ...Menu Summary for R S SMV03 R S SML R S SMV03 1090 3123 12 E 6 3 16 Menu Summary for R S SMV03 ...
Страница 148: ......
Страница 240: ......
Страница 242: ......