Test sequence
R&S®SML / R&S®SMV03
1090.3123.12 10.30
E-6
Stereo modulation (option R&S SML-B5)
Frequency response
Test setup
Test setup 6
Settings on R&S SML / R&S
SMV03
- PRESET
- FREQ 100MHz
- LEVEL 0 dBm
- MODULATION STEREO SOURCE LFGEN (or EXT L/R)
MODE L (or MODE R)
Setting on UPL
- GEN:
- INSTRUMENT ANALOG: Channel(s) 1 (or 2)
- FUNCTION Sine
- FUNCTION Frequency = 1kHz
- FUNCTION Voltage = 0.707V
Setting on FMB
- DEMODULATOR FM-STEREO
- CHANNEL L (or R)
- RELATIVE
Measurement
Set the respective LF generator (R&S SML or UPL) to 500Hz and
measure the reference level (MEAS-REF). Then traverse the LF
gen. from 20Hz to 15kHz and record the frequency response in all
four combinations (internal L, internal R, external L, external R).
Total harmonic distortion and channel separation
Test setup
Test setup 6
Settings on R&S SML / R&S
SMV03
- PRESET
- FREQ
Measurement frequency
Measurement frequency:
10.7MHz, 66 MHz, 76MHz, 87MHz, 98MHz, 110MHz
- LEVEL 0 dBm
- MODULATION STEREO SOURCE EXT L/R,
MODE L != R
- AF = 1kHz.
Setting on UPL
- FILE:
- LOAD INSTRUMENT STATE: Mode = Default
- GEN:
- INSTRUMENT ANALOG: Channel(s) 1 (or 2)
- FUNCTION: Voltage = 0.707 V
- ANLR:
- INSTRUMENT: Channel(s) 1 & 2
- INSTRUMENT: Ch1, Ch2 Imped = 600 Ohm
- FREQ / PHASE: Meas Time = Precision
- FUNCTION: RMS & S/N
- FUNCTION: Unit Ch1, Ch2 = dBr
- FUNCTION: Reference Value = STORE Ch1 (or Ch2)
Содержание SML01
Страница 4: ......
Страница 14: ......
Страница 22: ......
Страница 32: ......
Страница 34: ......
Страница 36: ......
Страница 38: ......
Страница 40: ......
Страница 42: ......
Страница 44: ......
Страница 56: ......
Страница 62: ......
Страница 77: ...R S SML R S SMV03 Menu Summary for R S SML 1090 3123 12 E 6 3 15 Menu Summary for R S SML ...
Страница 78: ...Menu Summary for R S SMV03 R S SML R S SMV03 1090 3123 12 E 6 3 16 Menu Summary for R S SMV03 ...
Страница 148: ......
Страница 240: ......
Страница 242: ......