Stereo Modulation (Option R&S SML-B5)
1090.3123.12 4.41
E-6
Set the group sequence, e.g.:
STEReo:DIRect
"GS=0A,14A"
Group 14A with variants 0 to 3 is now transmitted.
Create a new AF list for the EON, using method A:
STEReo:DIRect
"eon-afa=1234,N,87.6,87.7,87.8"
Create further AF lists for the EON, using method A:
STEReo:DIRect
"EON-AFA=1234,+,88.6,88.7,88.8"
Read the first AF list of the EON with STEReo:DIRect?
"eon-afa,1234,1"
Note:
Do not combine methods A and B for generating EON alternative frequency lists.
Create a new AF list for the EON, using method B:
STEReo:DIRect
"EON-AFB=1234,N,87.6,87.7,87.8"
87.6 = tuned frequency,
87.7 = mapped frequency 1(variant 5),
87.8 = mapped frequency 2 variant 6)
A maximum of five AF lists can be generated. For type A lists, max. 25 frequencies per list can be
specified, for type B lists, max. five frequencies per list.
Free Format Groups (FFGs)
In the user-definable groups 1A, 3A, 5A, 6A, 7A, 8A, 9A, 10A, 11A, 12A and 13A, any desired data can
be transmitted. Five bits of this data are transmitted in block B and 16 bits each in blocks C and D of the
specified group.
1. Define the data to be transmitted in group 1A:
STEReo:DIRect “1A=05,0000000000,1FFFFFFFFF“
Group 1A is now transmitted first with "0000000000" and then with "1FFFFFFFFF".
Each of the two data sequences is retransmitted five times, which is indicated by the information
"05".
2. Set the group sequence, e.g.:
STEReo:DIRect
"gs=0a,1a"
The defined data is now transmitted in group 1A.
Max. 20 different data sequences can be defined.
Transparent-Mode
The transparent mode allows the user to transmit freely definable binary data instead of the standard
RDS data. Blocks A to D of the RDS groups are used. This means that standard RDS data will no
longer be transmitted when transparent data is set. The binary data will be sent even if it constitutes no
valid or meaningful RDS data. The transmission of standard RDS data will not be resumed until the
transparent data is deleted.
Delete the transparent data and switch back to standard RDS data transmission with:
STEReo:DIRect
“TRANS=0“
Max. 20 different data sequences can be defined.
Содержание SML01
Страница 4: ......
Страница 14: ......
Страница 22: ......
Страница 32: ......
Страница 34: ......
Страница 36: ......
Страница 38: ......
Страница 40: ......
Страница 42: ......
Страница 44: ......
Страница 56: ......
Страница 62: ......
Страница 77: ...R S SML R S SMV03 Menu Summary for R S SML 1090 3123 12 E 6 3 15 Menu Summary for R S SML ...
Страница 78: ...Menu Summary for R S SMV03 R S SML R S SMV03 1090 3123 12 E 6 3 16 Menu Summary for R S SMV03 ...
Страница 148: ......
Страница 240: ......
Страница 242: ......