Operation
Power Supply NGMO
4.00 / 07-2004
4-7
Example 1
line 1:
•
"
AS
" means that currently the settings for channel A are displayed.
If you turn the output on by pressing the "
OUTP. on/off
" key the
display changes to "
AM
" indicating that this is now a measured
value taken from the output terminals of the selected power supply.
•
x.xxx
is the output voltage. If "
AS
" is displayed (output off)
x.xxx
is
the nominal voltage. If "
AM
" is displayed (output on)
x.xxx
is the
voltage measured on the sense lines of the output terminals.
•
y.yyy
is the current limit / output current. IF "
AS
" or "
S
" is display-
ed in front of
y.yyy
then
y.yyy
is the current limit setting. If "AM" is
displayed
y.yyy
is the actual current, measured on the force lines
of the output terminals.
line 2:
•
A:OFF
means that the output of power supply A is currently turned
off.
•
A:ON
means that the output of power supply A is currently turned
on.
•
AB:OFF
is displayed if function A/B24 "Common output on/off"
(see below Menu Items) is turned on and both outputs are off.
•
AB:ON
is displayed if function A/B24 "Common output on/off" (see
below Menu Items) is turned on and both outputs are on.
•
PROT
is displayed if the
OCP
key has been pressed so activating
the overcurrent protection function. In
OCP
active mode, the output
is switched off if the current on output lines exceeds the current li-
mit setting (y.yyy).
If
OCP
is turned off, the current is limited to the current limit setting.
If NGMO2 enters the current limit state "
ILIM
" is displayed at the
same location as "
PROT
".
"INH" is displayed at the "PROT" location when the output has
been turned off by a logical 1 applied to the INHIB_A pin on the
"
CONTROL I/O
" connector.
•
RDY
is displayed when the NGMO2 has finished sampling and
measurement data are ready to be transferred for analysis.
•
REM
is displayed when the NGMO2 has received remote com-
mands via the IEEE interface or serial interface. In remote mode
Содержание NGMO1
Страница 8: ...Power Supply NGMO Tables 4 Version 4 00 07 2004 ...
Страница 14: ...Power Supply NGMO Safety 2 2 4 00 07 2004 ...
Страница 51: ...Operation Power Supply NGMO 4 00 07 2004 4 29 Fig 4 7 Status model structure ...
Страница 52: ...Power Supply NGMO Operation 4 30 4 00 07 2004 ...
Страница 56: ...Power Supply NGMO IEEE Commands 5 4 4 00 07 2004 Com mand Parameter Par value Description Def SCPI MAX DVMB ...
Страница 84: ...Power Supply NGMO Ordering Information 6 2 4 00 07 2004 ...