Power Supply NGMO
Operation
4-26
4.00 / 07-2004
Fig. 4-6
Measurement repetition
NOTE:
The post- or pre-trigger is only available for a single sample
sequence (triggercount = 1). With a selected triggercount > 1 data
acquisition starts at the trigger signal for any sequence.
For each measurement range the trigger level is settable in the speci-
fied value borders with the given resolution of that range. With setting
the trigger level to the low border of the selected measurement range,
the AUTO TRIGGER is activated for that range. In this mode when
starting a sample sequence the trigger system searches for a signifi-
cant value change of the watched signal and sets the trigger level au-
tomatic if such a value change occurs.
In remote mode also software triggered measurments are possible with
the NGMO.
4.5.2.2 Pulse
Analysing
The NGMO2 provides the analysing of pulsed signals, which delivers
the values of several pulse characterising sizes. This values are formed
out of the samples as described below.
For all values the following applies to the result. With a selected trig-
gercount x greater than one at first the pulse value is formed for each of
the x single measurement sequences and than the average of these x
values is formed to get the final result.
Содержание NGMO1
Страница 8: ...Power Supply NGMO Tables 4 Version 4 00 07 2004 ...
Страница 14: ...Power Supply NGMO Safety 2 2 4 00 07 2004 ...
Страница 51: ...Operation Power Supply NGMO 4 00 07 2004 4 29 Fig 4 7 Status model structure ...
Страница 52: ...Power Supply NGMO Operation 4 30 4 00 07 2004 ...
Страница 56: ...Power Supply NGMO IEEE Commands 5 4 4 00 07 2004 Com mand Parameter Par value Description Def SCPI MAX DVMB ...
Страница 84: ...Power Supply NGMO Ordering Information 6 2 4 00 07 2004 ...