IEEE Commands
Power Supply NGMO
4.00 / 07-2004
5-13
Com-
mand
Parameter
Par. value
Description
Def.
SCPI
1E-4 <=
NUM_VAL <= 7
MIN
0.0
DEFault
2.5
MAX
5.0
:PULSe
:TRIGger
:LEVel
:LOW
<CHAR_VAL> or
<NUM_VAL>
Sets trigger level
for 0.5A (LOW)
range
2.5 mA
1.00E-05
0.01 mA
0.5
500 mA
AUTO or 0
Auto trigger
MIN
0A -> AUTO
*
DEFault
250mA
MAX
500 mA
DISPlay
Sets LCD to
TRIGGER LEVEL
RANGE 0.5 A
:PULSe
:TRIGger
:LEVel
:LOW?
<CHAR_VAL>
Queries trigger
level for 5mA
range
Results:
1E-5 <=
NUM_VAL <= 0.5
MIN
0.0
DEFault
2.5E-1
MAX
5.0E-1
:PULSe
:TRIGger
:LEVel
:DVM
<CHAR_VAL> or
<NUM_VAL>
Sets trigger level
DVM
0 V
-5.999
-5.999 Volt
25.000
25.000 Volt
MIN
AUTO
*
DEFault
0 Volt
MAX
25.000 Volt
DISPlay
Sets LCD to
TRIGGER LEVEL
DVM
AUTO
Auto trigger
:PULSe
:TRIGger
:LEVel
:DVM?
<CHAR_VAL>
Queries trigger
level DVM
Results:
-5.999 <=
NUM_VAL <=
25.000
MIN
AUTO
DEFault
0.000
MAX
25.000
:PULSe
:TRIGger
:SOURce
<CHAR_VAL>
Selects trigger
source
INT
INT
Intern
EXT
Extern
Содержание NGMO1
Страница 8: ...Power Supply NGMO Tables 4 Version 4 00 07 2004 ...
Страница 14: ...Power Supply NGMO Safety 2 2 4 00 07 2004 ...
Страница 51: ...Operation Power Supply NGMO 4 00 07 2004 4 29 Fig 4 7 Status model structure ...
Страница 52: ...Power Supply NGMO Operation 4 30 4 00 07 2004 ...
Страница 56: ...Power Supply NGMO IEEE Commands 5 4 4 00 07 2004 Com mand Parameter Par value Description Def SCPI MAX DVMB ...
Страница 84: ...Power Supply NGMO Ordering Information 6 2 4 00 07 2004 ...