Power Supply NGMO
Unit Description
3-6
4.00 / 07-2004
- Analysis of values
PEAK, MIN,
HIGH, LOW,
RMS,
AVERage
PEAK, MIN,
HIGH, LOW,
RMS,
AVERage
Protection Functions
- OVP
1.5 to 22 V, settable
1.5 to 22 V, settable
- OCP
0 to 5 A, settable
0 to 5 A, settable
- Open-sense lead detection
±
(0 to 4 V) about set
voltage, settable
±
(0 to 4 V) about set
voltage, settable
General Data
- Remote control
IEEE488.2, RS232C
- Control inputs
2 x measurement trigger, 2 x output inhibit
- Control outputs
2 x complete, 4 x relay driver, fault
- AC supply
110/115 V and 220/230 V selectable, 50 to 60 Hz
- AC input power
220 VA max.
220 VA max.
- Dimensions
210.8 mm x 87.6 mm x 420 mm without rear and buttom
feet
- Mass
7.5 kg
Environmental Data
- Humidity
85 % at 35 °C non-condensing
- Operating temperature range
0 to 45 °C
- Nominal temperature range
5 to 40 °C, 0 to 35 °C full power
- Storage temperature range
-40 to 70 °C
- Temperature coefficient
18 to 28 °C (full accuracy)
5 to 40 °C (0.1 x specification)/°C
Содержание NGMO1
Страница 8: ...Power Supply NGMO Tables 4 Version 4 00 07 2004 ...
Страница 14: ...Power Supply NGMO Safety 2 2 4 00 07 2004 ...
Страница 51: ...Operation Power Supply NGMO 4 00 07 2004 4 29 Fig 4 7 Status model structure ...
Страница 52: ...Power Supply NGMO Operation 4 30 4 00 07 2004 ...
Страница 56: ...Power Supply NGMO IEEE Commands 5 4 4 00 07 2004 Com mand Parameter Par value Description Def SCPI MAX DVMB ...