Operating with ARGUS-UMS 

UMS100/120/120 System Manual  

49 3030.3059.72-02.00 

Step 3: 

 
In Step 3 the measurement settings are selected.  

If stronger signals are expected, select -20 dB or even -30 dB at 

Gain Control

.

For very weak signals select the +25 dB setting. 
In case of frequently changing signal levels the 

A

utomatic 

G

ain 

C

ontrol (

Fast

, i.e. with FFM also 

Medium

/S

low

) may be the preferred choice.  

Keep the measurement period for scans as short as possible, select 

1 ms for fast scans.  

Demodulation is not important during a scan.  

For fixed frequency measurements also 

Frequency

 and 

Offset

 can be selected or - alternatively 

Audio

. In this case the demodulated audio signal will be directly recorded on the UMS100/120. For 

this purpose a demodulator must be switched on which is not possible at this point. Please refer to 
chapter 2.4.3. 

If there is an active connection to the UMS100/120 during AMM, the continuous measurements can 
then be shown in the form of a graph. You can select the graphic type here. 

 

Figure 2-18: AMM Wizard Step 3, Measurement Definition 

 

Step 4: 

 
In the last but one configuration window, set when and if measurement values are to be 

saved

 as well 

as if and how an 

alarm

 is to be triggered automatically at the Control PC. in the center. The <Alarm> 

button opens a window for this purpose. Here the essential things are now defined for automatic 
operation. 

If measurement results are to be saved on the UMS100/120 the 

Measurement Time

 preset must 

be taken into account. A continuously running measurement creates huge amounts of data and the 
memory runs at full capacity. Besides fast scans, this can occur particularly when there are 
continuous audio recordings (see Section 2.4.1). 

 

 
What may be more important is that the large amounts of recorded data must also be “moved” from 
UMS100/120 to the Control PC using an extremely slow interface. For example, a limited audio 
recording with RF level takes at least approx. one minute to transfer, even with a fast LAN 
connection. Using a GSM connection a transfer would take approx. 6 hours. 

Содержание 3030.3013.02

Страница 1: ...3030 3059 72 02 00 System Manual Division Radiomonitoring and Radiolocation Monitoring System R S UMS100 120 with ARGUS UMS Software 3030 3013 02 3035 1025K02 3034 0090 02...

Страница 2: ...gennamen sind Warenzeichen der jeweiligen Eigent mer R S is a registered trademark of Rohde Schwarz GmbH Co KG Proper names are trademarks of the respective owners Printed in the Federal Republic of G...

Страница 3: ...UMS100 120HF R S UMS100 120SHF is abbreviated with UMS100 120SHF R S ARGUS UMS is abbreviated with ARGUS UMS R S ARGUS is abbreviated with ARGUS R S is a registered trademark of Rohde Schwarz GmbH Co...

Страница 4: ......

Страница 5: ...2 2 3 Manual operation in Direct Measurement Mode DMM 35 2 3 1 Gain Control 36 2 3 2 Overload Indicator 36 2 3 3 Getting an overview of the spectrum using Scan 37 2 3 4 2D Waterfall Diagram in Scan Mo...

Страница 6: ...UMS100 120 120 System Manual Contents 3030 3059 72 02 00 4...

Страница 7: ...d connection 33 Figure 2 4 ARGUS UMS connected to a UMS100 120 34 Figure 2 5 Antenna information 35 Figure 2 6 Example of a fast scan for a wide frequency range 37 Figure 2 7 Example Scan for high res...

Страница 8: ...0 System Manual Figures 3030 3059 72 02 00 6 Figure 2 34 Analyzing and processing measurement data 62 Figure 2 35 Audio playback 63 Figure 3 1 Service Tool Firmware Update 64 Figure 3 2 Service Tool M...

Страница 9: ...k In addition steps are described required for setting up measurements which run automatically on the UMS100 120 The system operator is then only required if a UMS100 120 detects a deviation and has t...

Страница 10: ...rk the configuration must take place before it is connected to the network LAN WAN as all UMS100 120 are set with a factory default address 192 168 0 31 One such configuration table is attached in app...

Страница 11: ...minimal service work when up and running and for the initial configuration a user interface can be used which is web based and can be started on each UMS100 120 A web browser such as MS Internet Expl...

Страница 12: ...on Connect the PC to the UMS100 120 in one of the two following ways Connect the RJ45 socket on the PC to the corresponding socket on the bottom of the UMS100 120 to be configured directly using a cro...

Страница 13: ...f the site see Figure 1 3 Acknowledge this certificate by clicking on Yes UMS120 incorporates the modular version of the UMS100 complete system Users therefore can connect to the UMS120 a large range...

Страница 14: ...UMS100 120 120 System Manual Operation preparation 3030 3059 72 02 00 12 Then the following Login window appears Figure 1 4 Service Tool UMS100 Login window Figure 1 5 Service Tool UMS120 Login window...

Страница 15: ...Firmware components CAUTION For reasons of security a timer runs which triggers an automatic logout after 30 minutes In this case the Login window appears again automatically Now the actual configura...

Страница 16: ...orrelated or automated measurements AMMs cannot start because start time has passed already The time must be the actual time in this time zone and not UTC time Check the data again and click on Submit...

Страница 17: ...ata again and click on Submit to send CAUTION Make a detailed note of the new access data and store it in a safe place If you forget your details or they get lost access to the UMS100 120 may no longe...

Страница 18: ...s ARGUS_MT1 if the UMS100 120 is controlled by ARGUS UMS software for standard ARGUS this may be different When all the fields are filled in and the Connection Type LAN or Dial up Connection has been...

Страница 19: ...they can also be entered at a later date CAUTION If you click on Submit once you have entered a new IP Address Subnet Mask and Default Gateway only necessary with a router the data is sent and a conne...

Страница 20: ...Put the options CD into the drive of the Control PC After selection of Browse mark the CD drive and the options file in there Activate install And finally click on Submit Figure 1 11 Options for UMS10...

Страница 21: ...esponding transducer tables and connection cable types and lengths are considered in the ARGUS control SW Through this correct field strength levels can be displayed In case of overlapping frequency r...

Страница 22: ...dBMV A combination of unknown antenna and known R S antenna nevertheless only displays voltage levels Transducer tables from other antenna suppliers can not be entered Now the UMS100 120 has been set...

Страница 23: ...the GSM connection which is available once the SIM card has been inserted and the power supply is connected up But further preparation is necessary on the control PC for setting up dial up networking...

Страница 24: ...ll a separate outgoing dial up networking using the Assistant for New Connections for each UMS100 120 connected via GSM Open the Network Connections folder using the MS Windows desktop start button or...

Страница 25: ...Properties window by clicking it with the right mouse button a At the General tab deselect All devices dial the same number and select the Show symbol for a connection in the information section of th...

Страница 26: ...can see in the following Figure 1 15 and click on Create Figure 1 15 Creating a user for dial up Please close the following empty mask with Close For reasons of security you must specify group member...

Страница 27: ...ties Member of window In the following the membership Power Users is exclusively entered Click on Add In the following Select Groups window click on Advanced Figure 1 17 Group selection In the expande...

Страница 28: ...02 00 26 Figure 1 18 Group selection expanded Close the Select Users window with OK also The name Properties window now only has the entry Power Users Figure 1 19 name Properties with entry Close nam...

Страница 29: ...except the modem already configured for the outgoing connection Click on Next to open the next window In the next Incoming VPN connection window you must select VPN connections not permitted Click on...

Страница 30: ...gher than in the first A total of two IP addresses must be available as shown in Figure 1 21 Figure 1 21 Assigning an IP address for dial up networking Transfer the other settings from Figure 1 21 Cli...

Страница 31: ...still be made in System View tab in Configuration Control Unit Configuration command in the menu Now a new entry for a UMS100 120 can be generated In the Input Field area under Measurement PC overwrit...

Страница 32: ...ick on them The procedure is now complete for configuring a UMS100 120 Click on OK to close the window If you want to set up more UMS100 120 begin with the intermediate step each time and carry out th...

Страница 33: ...nal windows opened from this window can be placed on the desktop They are not restricted to the size of the standard window You can view information in the help text on the buttons and input fields by...

Страница 34: ...s Connect as well as Connect and Activate Depending on the type of connection setting up the UMS100 120 connection takes up to one minute During this time a corresponding window appears For dial up ne...

Страница 35: ...tenna configuration when options antenna configurations from the Service Tool were loaded or deleted the data area was emptied with the Clean Up Data Area maintenance function from the Service Tool in...

Страница 36: ...of lightening on a yellow background shows the active remote connection The input fields for the geographical coordinates and height can be filled in or changed These parameters are transferred to th...

Страница 37: ...ingle frequency measurement level frequency offset Scan with start and stop frequency input and increments for a quick overview Frequency List Scan FLS single frequency measurement for preset frequenc...

Страница 38: ...es a critical value where the pre amplifier could produce intermodulation or more critical the analogue digital converter could be overdriven the indicator color is changing from green as shown above...

Страница 39: ...h and the increment is important However you can overlook this ratio as the Detector is generally in a max peak position Each signal is then displayed even within larger increments The strongest signa...

Страница 40: ...her scan may be necessary over a more limited frequency range with smaller steps 2 3 4 Waterfall Diagram in Scan Mode The examples in chapter 2 3 3 show also the 2D Waterfall Diagram Here the signals...

Страница 41: ...own in the line above the scan graphic Figure 2 8 Scan with Max Hold 2 3 6 Listen Mode in Scan Before starting the scan a suitable Demodulation AM FM IF Bandwidth Audio On Audio Transf Rate and Volume...

Страница 42: ...is always performed with Max Peak Also the step width should be made small enough to get a fine resolution After starting the scan the center frequency marker is manually set with the mouse or by ent...

Страница 43: ...owing options can be selected Set Marker Frequency with switchover to FFM the marker frequency will be the measured frequency Transfer Settings to Single Mode The settings for IF Bandwidth Meas Time e...

Страница 44: ...rmed Directly after clicking the Start button the Level is displayed and the audio modulation can be heard if the relevant demodulator is selected and Audio is switched on If the measurement is not ru...

Страница 45: ...to the set frequency can be up to some kHz The above example Figure 2 12 shows the frequency of a narrow band FM signal where a measurement is realistic for such kind of transmitters The accuracy can...

Страница 46: ...UMS100 120 120 System Manual Operating with ARGUS UMS 3030 3059 72 02 00 44 Figure 2 13 Frequency and offset measurement for AM carrier...

Страница 47: ...ned and started first run a test with specific settings in the manual measurement mode There is a range of combinations which exclude one another In addition there are limits for measurement times ban...

Страница 48: ...rding 1 65 5 000 data records FFM at 105 7 MHz meas time 20 ms IF bandwidth 100 kHz pre amplifier one level measurement with continuous measurement audio recording with SpeexQ1 quality 0 43 4 970 data...

Страница 49: ...data is recorded over a longer time period and for example measurements are carried out only every 15 minutes so as to save memory space Note The total available memory space is not the issue here As...

Страница 50: ...to the type signal shall be monitored the Step Width is usually adjusted to the channel spacing of the radio service in that frequency range Do not follow this rule in case that high scan speeds over...

Страница 51: ...can then be shown in the form of a graph You can select the graphic type here Figure 2 18 AMM Wizard Step 3 Measurement Definition Step 4 In the last but one configuration window set when and if meas...

Страница 52: ...Message function and a Signal Tone are only released on the Control PC if there is an active connection In case SMS is selected and a mobile phone number with information text has been entered the UM...

Страница 53: ...tic alarming At the Station input field the name Control Unit is always shown in ARGUS UMS The Delay Time determines a delay after the first alarm has been detected before the UMS100 120 connects to t...

Страница 54: ...UMS100 120 120 System Manual Operating with ARGUS UMS 3030 3059 72 02 00 52 Figure 2 20 AMM window other functions...

Страница 55: ...can be extended and more specified here Several AMM at different times can be defined using different frequency ranges etc and inserting them into the table Figure 2 21 AMM Timing Definition window T...

Страница 56: ...closing a window if changes have been made before After closing a window a question appears if changes should be saved Always accept this question Figure 2 23 AMM Range Definition window In the Measu...

Страница 57: ...Operating with ARGUS UMS UMS100 120 120 System Manual 55 3030 3059 72 02 00 Figure 2 24 AMM Further Settings window...

Страница 58: ...General Log File Once the AMM has started the connection to the UMS100 120 can be interrupted in the Select window Alarms are still transferred and displayed in the General Log File The General Log Fi...

Страница 59: ...pends on whether the alarm goes off when the Limit Line is later overshoot or undershoot in the automatic measurement A positive offset of 5 dB for example pushes the limit up Slight level fluctuation...

Страница 60: ...ou want to scan After closing the list it directly can be used for measuring Figure 2 28 Creating a frequency list Instead of entering manually you can also import a frequency list created in Microsof...

Страница 61: ...Operating with ARGUS UMS UMS100 120 120 System Manual 59 3030 3059 72 02 00 Figure 2 29 Take over frequencies into the frequency list...

Страница 62: ...eneral Log File Figure 2 30 Disk capacity shortage on UMS100 120 Now the following tasks should be done Make a connection to the relevant UMS100 120 if there is not one already Open the Navigator and...

Страница 63: ...Operating with ARGUS UMS UMS100 120 120 System Manual 61 3030 3059 72 02 00 Figure 2 31 Measurement result files on UMS100 120 Figure 2 32 Corresponding File Info...

Страница 64: ...rtently copy very large files particularly audio which cannot be transferred within an acceptable time due to a slower GSM connection click on Delete to erase the files directly After transferring the...

Страница 65: ...dy have a customized user interface geared to their specific purposes which it is essential to keep For these situations there is a fitting solution by using Order Report Module with ARGUS Data Channe...

Страница 66: ...to update the version of the control software on the control PC before a Firmware update on the UMS100 120 is performed otherwise operation is no longer possible Before the Firmware update clarify wit...

Страница 67: ...l Go to Maintenance and click on Execute that comes after Cleanup Data Area CAUTION If Clean up Data Area has been executed all saved measurement results on the UMS100 120 are lost If the problem has...

Страница 68: ...rohde schwarz com or by phone under 49 89 4129 12194 Further information may be available on the following Web Site www ums rohde schwarz com For a comprehensive service or for hardware faults the UM...

Страница 69: ...ile network The configuration that is not required is in gray It is advisable to enter these settings beforehand Otherwise a service done locally on the LAN interface of a UMS100 connected by GSM must...

Страница 70: ......

Отзывы: