Roche Diagnostics
B-32
Operator’s Manual · Version 1.0
4 Cassette
URISYS 2400
First-time installation
18
Press <Home> to return to the [Routine Monitor] screen. The screen
shows the number of remaining strips as 400, followed by the date the
cassette was installed.
Handling precautions
When handling test strips, the following precautions must be observed:
o
Only use a URISYS 2400 cassette that has not exceeded its expiration date.
o
Store cassette packages at a temperature between 2-30°C (36-86°F).
o
Refrigerated cassettes must be at room temperature for at least two hours
before opening the package and using the cassette.
o
Check the package for damage.
o
After removing the cassette from the packaging, ensure the test strip layers
are correctly aligned and the test pads are not unusually discolored.
o
Do not directly touch the test pads of the test strip.
o
Do not let the strips come in contact with water or with any volatile,
strongly acidic, or alkaline chemicals.
o
Install the cassette into the analyzer within 3 minutes of opening the seal.
o
Keep the cassette in the firmly locked cassette compartment until the test
strips are completely used up.
o
Do not open the cassette compartment door as long as the cassette is not
empty or the on-board stability has not expired.
o
Follow the instructions provided with the URISYS 2400 cassette by the
manufacturer and read carefully all information on the package insert.
When a cassette becomes empty during sample processing, replacement by a new
cassette can be performed through the [Routine Monitor] screen by activating the
<Change Cassette> button, (see “Replace an empty URISYS 2400 cassette” on
page C-18).
Содержание URISYS 2400
Страница 1: ...Operator s Manual Version 1 0 URISYS 2400...
Страница 4: ...Roche Diagnostics iv Operator s Manual Version 1 0 URISYS 2400...
Страница 6: ...URISYS 2400 Roche Diagnostics vi Operator s Manual Version 1 0...
Страница 14: ...Roche Diagnostics xiv Operator s Manual Version 1 0 URISYS 2400...
Страница 15: ...Part A is an overview of the URISYS 2400 Overview A...
Страница 16: ......
Страница 23: ...Part B is the system description of the URISYS 2400 System description B...
Страница 24: ......
Страница 57: ...Part C contains the operating procedures for the URISYS 2400 Operation C...
Страница 58: ......
Страница 102: ...Roche Diagnostics C 44 Operator s Manual Version 1 0 6 Special operations URISYS 2400 Checks...
Страница 103: ...Part D describes all maintenance procedures relating to the URISYS 2400 Maintenance D...
Страница 104: ......
Страница 121: ...Part E lists and describes the meaning of the alarm flags which are shown by the URISYS 2400 Troubleshooting E...
Страница 122: ......
Страница 132: ...Roche Diagnostics E 10 Operator s Manual Version 1 0 8 Data alarms flags URISYS 2400 Reset the analyzer...
Страница 146: ...Roche Diagnostics E 24 Operator s Manual Version 1 0 9 Instrument alarms messages URISYS 2400 Instrument alarms messages...
Страница 147: ...Part F describes the technical specifications relating to the URISYS 2400 Appendix F...
Страница 148: ......
Страница 155: ...Glossary and index G...
Страница 156: ......
Страница 166: ...Index URISYS 2400 Test parameters Wavelengths Roche Diagnostics G 10 Operator s Manual Version 1 0...