Roche Diagnostics
Operator’s Manual · Version 1.0
E-19
URISYS 2400
9 Instrument alarms (messages)
Instrument alarms (messages)
45
Defragment Memory
Flash Memory Error
1
STOP
1. Switch off the analyzer, then restart the
analyzer.
2. If the alarm re-occurs, create a backup
file of the alarm trace data on a blank
diskette by executing the FD write
procedure on the [Alarm Trace] screen.
3. Contact Technical Support.
46
Data Area full
Flash Memory Error
1-3
S. STOP
1. Remaining capacity of the routine
sample data storage is 16 or less
(subcategory 1) at Start.
2. STAT sample data storage is six or less
(subcategory 2) at Start.
3. Control sample data storage is 10 or
less (subcategory 3) at Start.
4-6
WARNING
1. Remaining capacity of the routine
sample data storage is 45 or less
(subcategory 4) at Start.
2. STAT sample data storage is 27 or less
(subcategory 6) at Start.
3. Execute the data delete procedure after
backing up the old sample data to a
diskette.
47
FLASH Memory File
open
Flash Memory Error
1-9
STOP
1. Switch off the analyzer, then restart the
analyzer.
2. If the alarm re-occurs, create a backup
file of the alarm trace data on a blank
diskette by executing the FD write
procedure on the [Alarm Trace] screen.
3. Contact Technical Support.
48
FLASH Memory File
read
Flash Memory Error
1-9
STOP
1. Switch off the analyzer, then restart the
analyzer.
2. If the alarm re-occurs, create a backup
file of the alarm trace data on a blank
diskette by executing the FD write
procedure on the [Alarm Trace] screen.
3. Contact Technical Support.
Category
Alarm name
Type of error
Subcategory
Level
Check and countermeasure
Table E-3
Instrument Alarms (messages) (continued)
Содержание URISYS 2400
Страница 1: ...Operator s Manual Version 1 0 URISYS 2400...
Страница 4: ...Roche Diagnostics iv Operator s Manual Version 1 0 URISYS 2400...
Страница 6: ...URISYS 2400 Roche Diagnostics vi Operator s Manual Version 1 0...
Страница 14: ...Roche Diagnostics xiv Operator s Manual Version 1 0 URISYS 2400...
Страница 15: ...Part A is an overview of the URISYS 2400 Overview A...
Страница 16: ......
Страница 23: ...Part B is the system description of the URISYS 2400 System description B...
Страница 24: ......
Страница 57: ...Part C contains the operating procedures for the URISYS 2400 Operation C...
Страница 58: ......
Страница 102: ...Roche Diagnostics C 44 Operator s Manual Version 1 0 6 Special operations URISYS 2400 Checks...
Страница 103: ...Part D describes all maintenance procedures relating to the URISYS 2400 Maintenance D...
Страница 104: ......
Страница 121: ...Part E lists and describes the meaning of the alarm flags which are shown by the URISYS 2400 Troubleshooting E...
Страница 122: ......
Страница 132: ...Roche Diagnostics E 10 Operator s Manual Version 1 0 8 Data alarms flags URISYS 2400 Reset the analyzer...
Страница 146: ...Roche Diagnostics E 24 Operator s Manual Version 1 0 9 Instrument alarms messages URISYS 2400 Instrument alarms messages...
Страница 147: ...Part F describes the technical specifications relating to the URISYS 2400 Appendix F...
Страница 148: ......
Страница 155: ...Glossary and index G...
Страница 156: ......
Страница 166: ...Index URISYS 2400 Test parameters Wavelengths Roche Diagnostics G 10 Operator s Manual Version 1 0...