9 Instrument alarms (messages)
URISYS 2400
Instrument alarms (messages)
12
DC Power
DC Power Supply Error
1-2
STOP
1. Contact Technical Support.
3
E.STOP
1. Contact Technical Support.
13
Water Supply Tank
Water Sensor Error
1
STOP
1. Check that the cable is connected.
Switch off the power.
2. Attach the connector to the water
supply tank.
3. Switch on the power.
14
Short of water
No Water Error
1
S.STOP
STOP
1. Check the float switch.
2. Add rinsing water to the water supply
tank, and try to restart.
2
WARNING
1. Check the float switch.
2. Add rinsing water to the water supply
tank, and try to restart.
15
Liquid Waste Tank
Waste Sensor Error
1
STOP
1. Check that the cable is connected.
2. Switch off the power.
3. Attach the connector to the liquid
waste tank.
4. Switch on the power.
16
Liquid Waste Tank
full
Full Waste Error
1
S.STOP
1. Empty the liquid waste tank.
17
Watch Dog Timer
Timing Error
1
E.STOP
1. Create a backup file of alarm trace data
on a new diskette, by executing the FD
write procedure on the [Alarm Trace]
screen.
2. Contact Technical Support.
19
TS Cassette empty
No Test Strip
1
S.STOP
1. Replace the URISYS 2400 cassette.
2
WARNING
1. Test Strip number below alarm level
3
WARNING
1. Test Strip number below alarm level
20
STAT Position
STAT Position Error
1
WARNING
1. Place the STAT sample at the STAT
sampling position, and request a STAT
measurement.
3
WARNING
1. Place the washing solution at the STAT
sampling position, and perform the
wash procedure.
Category
Alarm name
Type of error
Subcategory
Level
Check and countermeasure
Table E-3
Instrument Alarms (messages) (continued)
Содержание URISYS 2400
Страница 1: ...Operator s Manual Version 1 0 URISYS 2400...
Страница 4: ...Roche Diagnostics iv Operator s Manual Version 1 0 URISYS 2400...
Страница 6: ...URISYS 2400 Roche Diagnostics vi Operator s Manual Version 1 0...
Страница 14: ...Roche Diagnostics xiv Operator s Manual Version 1 0 URISYS 2400...
Страница 15: ...Part A is an overview of the URISYS 2400 Overview A...
Страница 16: ......
Страница 23: ...Part B is the system description of the URISYS 2400 System description B...
Страница 24: ......
Страница 57: ...Part C contains the operating procedures for the URISYS 2400 Operation C...
Страница 58: ......
Страница 102: ...Roche Diagnostics C 44 Operator s Manual Version 1 0 6 Special operations URISYS 2400 Checks...
Страница 103: ...Part D describes all maintenance procedures relating to the URISYS 2400 Maintenance D...
Страница 104: ......
Страница 121: ...Part E lists and describes the meaning of the alarm flags which are shown by the URISYS 2400 Troubleshooting E...
Страница 122: ......
Страница 132: ...Roche Diagnostics E 10 Operator s Manual Version 1 0 8 Data alarms flags URISYS 2400 Reset the analyzer...
Страница 146: ...Roche Diagnostics E 24 Operator s Manual Version 1 0 9 Instrument alarms messages URISYS 2400 Instrument alarms messages...
Страница 147: ...Part F describes the technical specifications relating to the URISYS 2400 Appendix F...
Страница 148: ......
Страница 155: ...Glossary and index G...
Страница 156: ......
Страница 166: ...Index URISYS 2400 Test parameters Wavelengths Roche Diagnostics G 10 Operator s Manual Version 1 0...