Roche Diagnostics
Operator’s Manual · Version 1.0
D-5
URISYS 2400
7 General maintenance actions
Maintenance menu
a
To use the washing solution at the STAT position
1
Set the analyzer to Stand-By mode.
2
Place a sample tube containing 10 ml of washing solution at position 1 of a
sample rack, and set the rack in the STAT position.
3
Press <Menu> on the [Routine Monitor] screen. The [Menu] screen is
displayed.
4
Press <Maintenance>. The [Maintenance] screen is displayed.
5
Press <Wash>. The [Wash] window appears.
6
Press <Execute> to start the wash procedure. A second [Wash] window
appears with the message “Executing”.
When handling the washing solution, ensure that you wear gloves and safety
goggles to avoid direct contact with the solution. If the solution contacts the body,
wash immediately with a large volume of water. Consult your laboratory protocol for
handling biohazard materials. If a sample is spilled on the analyzer, wipe up
immediately and apply disinfectant.
Do not remove the rack with the washing solution until the wash procedure is
finished.
If the washing cycle is interrupted, carry out the wash procedure again from the
beginning. Otherwise, the remaining detergent in the flow path may damage the
instrument or cause incorrect data.
When the <Stop> button on the [Executing] window is pressed, a [Warning] window
alerts the operator to the risks of interrupting the washing process.
Содержание URISYS 2400
Страница 1: ...Operator s Manual Version 1 0 URISYS 2400...
Страница 4: ...Roche Diagnostics iv Operator s Manual Version 1 0 URISYS 2400...
Страница 6: ...URISYS 2400 Roche Diagnostics vi Operator s Manual Version 1 0...
Страница 14: ...Roche Diagnostics xiv Operator s Manual Version 1 0 URISYS 2400...
Страница 15: ...Part A is an overview of the URISYS 2400 Overview A...
Страница 16: ......
Страница 23: ...Part B is the system description of the URISYS 2400 System description B...
Страница 24: ......
Страница 57: ...Part C contains the operating procedures for the URISYS 2400 Operation C...
Страница 58: ......
Страница 102: ...Roche Diagnostics C 44 Operator s Manual Version 1 0 6 Special operations URISYS 2400 Checks...
Страница 103: ...Part D describes all maintenance procedures relating to the URISYS 2400 Maintenance D...
Страница 104: ......
Страница 121: ...Part E lists and describes the meaning of the alarm flags which are shown by the URISYS 2400 Troubleshooting E...
Страница 122: ......
Страница 132: ...Roche Diagnostics E 10 Operator s Manual Version 1 0 8 Data alarms flags URISYS 2400 Reset the analyzer...
Страница 146: ...Roche Diagnostics E 24 Operator s Manual Version 1 0 9 Instrument alarms messages URISYS 2400 Instrument alarms messages...
Страница 147: ...Part F describes the technical specifications relating to the URISYS 2400 Appendix F...
Страница 148: ......
Страница 155: ...Glossary and index G...
Страница 156: ......
Страница 166: ...Index URISYS 2400 Test parameters Wavelengths Roche Diagnostics G 10 Operator s Manual Version 1 0...