Roche Diagnostics
Operator’s Manual · Version 1.0
E-15
URISYS 2400
9 Instrument alarms (messages)
Instrument alarms (messages)
21
Motor Controller
Motor Error
1-17
STOP
1. Contact Technical Support.
22
Motor Time-out
Motor Error
1-17
STOP
1. Contact Technical Support.
23
Sample Short
Sample Volume Error
1
WARNING
1. Probe does not detect sufficient
sample. Add more sample, and
measure again.
2. If there is sufficient sample volume, and
there is still an alarm, clean the probe
assembly.
2
STOP
1. Probe does not detect sufficient
sample. Add more sample, and
measure again. If there is sufficient
sample volume, and there is still an
alarm, clean the probe assembly.
24
No more racks
Last Rack Processed
1
WARNING
1. Processing of the last rack has been
completed, and there are no more racks
on the input line.
25
Rack Barcode
Rack Barcode Label
Error
1
S.STOP
1. Check whether the barcode label is
dirty.
2. Check whether the barcode is
positioned correctly.
3. Check whether the barcode label is
missing.
4. Check whether the barcode label has
the correct type of code.
5. Contact Technical Support.
2
S.STOP
1. The rack barcode is outside the
specified range.
2. Replace the barcode.
26
Rack Barcode
Reader
Rack Barcode Reading
Error
1
S. STOP
1. Check whether the barcode label is
dirty.
2. Check whether the barcode is
positioned correctly.
3. Check whether the barcode label is
missing.
4. Check whether the barcode label has
the correct type of code.
5. Contact Technical Support.
Category
Alarm name
Type of error
Subcategory
Level
Check and countermeasure
Table E-3
Instrument Alarms (messages) (continued)
Содержание URISYS 2400
Страница 1: ...Operator s Manual Version 1 0 URISYS 2400...
Страница 4: ...Roche Diagnostics iv Operator s Manual Version 1 0 URISYS 2400...
Страница 6: ...URISYS 2400 Roche Diagnostics vi Operator s Manual Version 1 0...
Страница 14: ...Roche Diagnostics xiv Operator s Manual Version 1 0 URISYS 2400...
Страница 15: ...Part A is an overview of the URISYS 2400 Overview A...
Страница 16: ......
Страница 23: ...Part B is the system description of the URISYS 2400 System description B...
Страница 24: ......
Страница 57: ...Part C contains the operating procedures for the URISYS 2400 Operation C...
Страница 58: ......
Страница 102: ...Roche Diagnostics C 44 Operator s Manual Version 1 0 6 Special operations URISYS 2400 Checks...
Страница 103: ...Part D describes all maintenance procedures relating to the URISYS 2400 Maintenance D...
Страница 104: ......
Страница 121: ...Part E lists and describes the meaning of the alarm flags which are shown by the URISYS 2400 Troubleshooting E...
Страница 122: ......
Страница 132: ...Roche Diagnostics E 10 Operator s Manual Version 1 0 8 Data alarms flags URISYS 2400 Reset the analyzer...
Страница 146: ...Roche Diagnostics E 24 Operator s Manual Version 1 0 9 Instrument alarms messages URISYS 2400 Instrument alarms messages...
Страница 147: ...Part F describes the technical specifications relating to the URISYS 2400 Appendix F...
Страница 148: ......
Страница 155: ...Glossary and index G...
Страница 156: ......
Страница 166: ...Index URISYS 2400 Test parameters Wavelengths Roche Diagnostics G 10 Operator s Manual Version 1 0...