8 Data alarms (flags)
URISYS 2400
Handle alarms
a
To download alarm lists from the analyzer to a diskette
1
Press <Menu> on the [Routine Monitor] screen. The [Routine Monitor]
screen is displayed.
2
Press <Tools>. The [Tools] screen is displayed.
3
Press <Alarm Trace>. The [Alarm Trace] screen is displayed.
4
Place diskette into diskette drive.
5
Press <FD write>. The [FD write] window appears.
6
Press <Execute>.
a
To clear alarms
1
Press <Menu> on the [Routine Monitor] screen. The [Routine Monitor]
screen is displayed.
2
Press <Tools>. The [Tools] screen is displayed.
3
Press <Alarm Trace>. The [Alarm Trace] screen is displayed.
4
Check the message.
5
Press <Clear Alarm>. The [Clear Alarm List] window appears.
6
Press <Cumu.> to clear the cumulative alarm list.
7
Press <Daily> to clear the daily alarm list.
o
The maximum number of daily alarms that can be stored is 255 logs or 10 types.
The maximum number of cumulative alarms that can be stored is 255 logs or 20
types.
o
Switching the power supply on or off is counted once.
o
When the date is changed or the power supply is switched off, the daily alarm
log is cleared.
Содержание URISYS 2400
Страница 1: ...Operator s Manual Version 1 0 URISYS 2400...
Страница 4: ...Roche Diagnostics iv Operator s Manual Version 1 0 URISYS 2400...
Страница 6: ...URISYS 2400 Roche Diagnostics vi Operator s Manual Version 1 0...
Страница 14: ...Roche Diagnostics xiv Operator s Manual Version 1 0 URISYS 2400...
Страница 15: ...Part A is an overview of the URISYS 2400 Overview A...
Страница 16: ......
Страница 23: ...Part B is the system description of the URISYS 2400 System description B...
Страница 24: ......
Страница 57: ...Part C contains the operating procedures for the URISYS 2400 Operation C...
Страница 58: ......
Страница 102: ...Roche Diagnostics C 44 Operator s Manual Version 1 0 6 Special operations URISYS 2400 Checks...
Страница 103: ...Part D describes all maintenance procedures relating to the URISYS 2400 Maintenance D...
Страница 104: ......
Страница 121: ...Part E lists and describes the meaning of the alarm flags which are shown by the URISYS 2400 Troubleshooting E...
Страница 122: ......
Страница 132: ...Roche Diagnostics E 10 Operator s Manual Version 1 0 8 Data alarms flags URISYS 2400 Reset the analyzer...
Страница 146: ...Roche Diagnostics E 24 Operator s Manual Version 1 0 9 Instrument alarms messages URISYS 2400 Instrument alarms messages...
Страница 147: ...Part F describes the technical specifications relating to the URISYS 2400 Appendix F...
Страница 148: ......
Страница 155: ...Glossary and index G...
Страница 156: ......
Страница 166: ...Index URISYS 2400 Test parameters Wavelengths Roche Diagnostics G 10 Operator s Manual Version 1 0...