Roche Diagnostics
E-16
Operator’s Manual · Version 1.0
9 Instrument alarms (messages)
URISYS 2400
Instrument alarms (messages)
27
Sample Barcode
Sample Barcode Label
Error
1-5
WARNING
1. Check whether the barcode label is
dirty.
2. Check whether the barcode is
positioned correctly.
3. Check whether the barcode label is
missing.
4. Check whether the barcode label has
the correct type of code.
5. Contact Technical Support.
28
Barcode Reader
Comm.
Barcode Reader Error
1-30
S.STOP
1. Check whether the barcode label has
the correct type of code.
2. Contact Technical Support.
29
Reflectance ADC
ADC Reflectance Error
1
WARNING
1. Contact Technical Support.
2-7
STOP
1. Contact Technical Support.
30
SG ADC
ADC Specific Gravity
Error
1
WARNING
1. Contact Technical Support.
2-7
STOP
1. Contact Technical Support.
31
CLA ADC
ADC Clarity Error
1
WARNING
1. Contact Technical Support.
2-7
STOP
1. Contact Technical Support.
33
Wavelength Setting?
Wavelength Error
2-10, 12
WARNING
1. Correctly certify the wavelength and
correction value with the blank pad.
2. The subcategory codes represent the
items listed as follows:
Subcategory Item
2pH
3LEU
4NIT
5PRO
6GLU
7KET
8UBG
9BIL
10ERY
11COL
12CLA
Category
Alarm name
Type of error
Subcategory
Level
Check and countermeasure
Table E-3
Instrument Alarms (messages) (continued)
Содержание URISYS 2400
Страница 1: ...Operator s Manual Version 1 0 URISYS 2400...
Страница 4: ...Roche Diagnostics iv Operator s Manual Version 1 0 URISYS 2400...
Страница 6: ...URISYS 2400 Roche Diagnostics vi Operator s Manual Version 1 0...
Страница 14: ...Roche Diagnostics xiv Operator s Manual Version 1 0 URISYS 2400...
Страница 15: ...Part A is an overview of the URISYS 2400 Overview A...
Страница 16: ......
Страница 23: ...Part B is the system description of the URISYS 2400 System description B...
Страница 24: ......
Страница 57: ...Part C contains the operating procedures for the URISYS 2400 Operation C...
Страница 58: ......
Страница 102: ...Roche Diagnostics C 44 Operator s Manual Version 1 0 6 Special operations URISYS 2400 Checks...
Страница 103: ...Part D describes all maintenance procedures relating to the URISYS 2400 Maintenance D...
Страница 104: ......
Страница 121: ...Part E lists and describes the meaning of the alarm flags which are shown by the URISYS 2400 Troubleshooting E...
Страница 122: ......
Страница 132: ...Roche Diagnostics E 10 Operator s Manual Version 1 0 8 Data alarms flags URISYS 2400 Reset the analyzer...
Страница 146: ...Roche Diagnostics E 24 Operator s Manual Version 1 0 9 Instrument alarms messages URISYS 2400 Instrument alarms messages...
Страница 147: ...Part F describes the technical specifications relating to the URISYS 2400 Appendix F...
Страница 148: ......
Страница 155: ...Glossary and index G...
Страница 156: ......
Страница 166: ...Index URISYS 2400 Test parameters Wavelengths Roche Diagnostics G 10 Operator s Manual Version 1 0...