8 Data alarms (flags)
Handle alarms
Handle alarms
The [Alarm monitor] screen displays up to four alarms.
When more than four messages are generated, press <Alarm Trace> on the
[Tools] screen to print the alarm codes and the times at which they occurred.
When the analyzer is switched on and in Stand-by status, any displayed alarm
must be checked, solved and cleared before next samples are processed.
If any errors have occurred during the setting of parameters, the message does
not appear until <Start> is pressed.
If the analyzer comes to an emergency stop due to an error, execute the reset
function on the [Maintenance] screen after resolving the alarm cause.
Alarm levels
The main alarm levels are WARNING, S.STOP, STOP, and E.STOP.
WARNING
Although this warning is issued, the analyzer continues to measure samples.
After completion of the current sampling, check the alarm condition, and take
the action outlined.
S.STOP
Only the sampling sequence is stopped. For samples in process, measurement
continues and the result data are generated. Check the alarm condition, take
the action outlined, and restart the operation.
STOP
The analyzer stops at the end of its current, 15-second instrument cycle. For
samples in process, measurement, result data printing and host
communication stop immediately. Check the alarm condition, and take the
action outlined. Check the results for the last samples that were processed
before the alarm message appeared. Reset all samples that were processed but
not measured/ printed/ sent to host and repeat the processing of these
samples.
E.STOP
The analyzer stops instantly (15-second instrument cycle is NOT finished).
For samples in process, measurement, result data printing and host
communication stop immediately. Check the alarm condition and take the
action outlined. Execute the reset function on the [Maintenance] screen after
resolving the alarm cause. Check the results for the last samples that were
processed before the alarm message appeared. Reset all samples that were
processed but not measured/ printed/ sent to host and repeat processing of
these samples.
Содержание URISYS 2400
Страница 1: ...Operator s Manual Version 1 0 URISYS 2400...
Страница 4: ...Roche Diagnostics iv Operator s Manual Version 1 0 URISYS 2400...
Страница 6: ...URISYS 2400 Roche Diagnostics vi Operator s Manual Version 1 0...
Страница 14: ...Roche Diagnostics xiv Operator s Manual Version 1 0 URISYS 2400...
Страница 15: ...Part A is an overview of the URISYS 2400 Overview A...
Страница 16: ......
Страница 23: ...Part B is the system description of the URISYS 2400 System description B...
Страница 24: ......
Страница 57: ...Part C contains the operating procedures for the URISYS 2400 Operation C...
Страница 58: ......
Страница 102: ...Roche Diagnostics C 44 Operator s Manual Version 1 0 6 Special operations URISYS 2400 Checks...
Страница 103: ...Part D describes all maintenance procedures relating to the URISYS 2400 Maintenance D...
Страница 104: ......
Страница 121: ...Part E lists and describes the meaning of the alarm flags which are shown by the URISYS 2400 Troubleshooting E...
Страница 122: ......
Страница 132: ...Roche Diagnostics E 10 Operator s Manual Version 1 0 8 Data alarms flags URISYS 2400 Reset the analyzer...
Страница 146: ...Roche Diagnostics E 24 Operator s Manual Version 1 0 9 Instrument alarms messages URISYS 2400 Instrument alarms messages...
Страница 147: ...Part F describes the technical specifications relating to the URISYS 2400 Appendix F...
Страница 148: ......
Страница 155: ...Glossary and index G...
Страница 156: ......
Страница 166: ...Index URISYS 2400 Test parameters Wavelengths Roche Diagnostics G 10 Operator s Manual Version 1 0...