14 • Setting the Bump Test and Calibration Parameters
Air Sample Time (Air)
The air sample time can be set separately for bump testing and
calibration. It is the length of time that a SM-2009U will draw air through
the AIR fitting on the back of the station. Air is drawn during a bump test
before an air adjust operation and to purge calibration gas from the
system after calibration gas has been drawn through the CAL. GAS fitting
on the back of the station.
Calibration Gas Sample Time (GAS)
The calibration gas sample time can be set separately for bump testing
and calibration. It is the length of time that a SM-2009U will draw
Table 3: Bump Test & Calibration Parameter
Parameter
Display
Tag
Available
Choices
Bump
Factory
Setting
Cal
Factory
Setting
Air Sample
Time
Air
• 30 seconds
• 45 seconds
• 60 seconds
• 90 seconds
30 seconds
45 seconds
Calibration Gas
Sample Time
GAS
• 30 seconds
• 45 seconds
• 60 seconds
• 90 seconds
• 120 seconds
30 seconds
90 seconds
Bump Test
Check Toler-
ance
cHE
• ± 10%
• ± 20%
• ± 30%
• ± 40%
• ± 50%
50%
n/a
Automatic Cali-
bration
cAL
• On
• Off
Off
n/a