![Rigol MSO8000 Series Скачать руководство пользователя страница 293](http://html1.mh-extra.com/html/rigol/mso8000-series/mso8000-series_user-manual_1461151293.webp)
RIGOL
MSO8000 User Guide
14-3
To Create a Mask
Press
Mask
→
Create
to self-define the mask of the pass/fail test. Press
Range
and
rotate the multifunction knob
to select the mask range. Press down the knob to
select it. You can also press
Range
continuously or enable the touch screen to select
it. You can select "Screen" or "Cursor" as the mask region. The default is "Screen".
⚫
If "Screen" is selected under
Range
, then the whole waveform display area is
considered to be the mask region. Press
X Mask
and
Y Mask
respectively.
Rotate the multifunction knob
at a small step or use the pop-up numeric
keypad to adjust the horizontal tolerance range and vertical tolerance range of
the test mask. During the adjustment, two white curves would be displayed in
the interface to show the outline of the current mask. Press
Create
to apply the
currently created mask (the region not covered by blue within the screen).
⚫
When "Cursor" is selected, two cursors (used to specify the current test mask
range) will be displayed on the screen. Cursor A is at the left and Cursor B is at
the right. Press
CursorA
and
CursorB
, and rotate the multifunction knob
to adjust the positions of both Cursor A and Cursor B, respectively. Also you can
press
CursorAB
and rotate the multifunction knob
to adjust the two
cursors synchronously (the spacing between Cursor A and Cursor B remains
unchanged). Then, refer to the above methods to adjust the horizontal and
vertical tolerance ranges, as well as create a mask (the region not covered by
blue within the cursor region).
To Save a Mask
When the pass/fail test function is enabled, you can save the current test mask range
to the internal memory or external USB storage device (when detected) in "*.pf"
format.
Press
Mask
→
Save
to enter the file saving interface. Please refer to descriptions in
to save the test mask file to the internal or external memory.
To Load a Mask
When the pass/fail test function is enabled, you can load the test mask files from the
internal memory or external USB storage device (when detected) and apply them to
the current pass/fail test function.
Press
Mask
→
Load
to enter the file loading interface. Please refer to descriptions in
to load the specified test mask files (in *.pf format) and apply
Содержание MSO8000 Series
Страница 1: ...User Guide MSO8000 Series Digital Oscilloscope Mar 2021 RIGOL TECHNOLOGIES CO LTD...
Страница 2: ......
Страница 24: ......
Страница 72: ......
Страница 78: ......
Страница 148: ......
Страница 202: ......
Страница 208: ......
Страница 214: ......
Страница 248: ...RIGOL Chapter 12 Protocol Decoding 12 14 MSO8000 User Guide...
Страница 264: ...RIGOL Chapter 12 Protocol Decoding 12 30 MSO8000 User Guide Sync Error...
Страница 274: ...RIGOL Chapter 12 Protocol Decoding 12 40 MSO8000 User Guide TSS Sync Frame Frame ID PL HCRC CYC Data TCRC...
Страница 284: ......
Страница 290: ......
Страница 296: ......
Страница 302: ......
Страница 308: ......
Страница 360: ......
Страница 383: ...Chapter 21 Remote Control RIGOL MSO8000 User Guide 21 3 5 6 3 4...
Страница 394: ......