Chapter 5 To Trigger the Oscilloscope
RIGOL
MSO8000 User Guide
5-45
Trigger Condition:
Press
When
, and rotate the multifunction knob
to select the desired trigger
condition, then press down the knob to select it. You can also press
When
continuously or enable the touch screen to select it.
⚫
Post: triggers on the specified position of the FlexRay bus. Press
Post
to select
"TSS End", "FSS_BSS End", "FES End", or "DTS End".
⚫
Frame: triggers on the frame of FlexRay bus.
➢
Press
Frame
to select the frame type. The types of frames include null,
Sync, Start, and All.
➢
Press
More
→
Define
to select "ID" or "Cyc Count".
When you select "ID", set the following parameters: ID Comp, ID Min, and
ID Max. Press
More
→
ID Comp
to select the comparison conditions. The
available choices include =,
≠
, >, <, ><, <>. Press
More
→
ID Max
or
press
More
→
ID Min
, then rotate the multifunction knob
or use the
pop-up numeric keypad to set the frame ID.
When you select "Cyc Count", set the following parameters: Cyc Comp,
Count Min, and Count Max. Press
More
→
Cyc Comp
to select the
comparison conditions. The available choices include =,
≠
, >, <, ><, <>.
Press
More
→
Count Max
or press
More
→
Count Min
, then rotate the
multifunction knob
or use the pop-up numeric keypad to set the cycle
count.
⚫
Symbol: triggers on the CAS/MTS (Collision Avoidance Symbol/Media Access
Test Symbol) and WUS (Wake Up Symbol) of FlexRay bus.
➢
Press
Symbol
to select the symbol type. The symbol type includes
CAS/MTS and WUS.
➢
Press
More
→
ID Comp
to select the comparison conditions. The available
choices include =,
≠
, >, <, ><, <>. Press
More
→
ID Max
or press
More
→
ID Min
, then rotate the multifunction knob
or use the pop-up
numeric keypad to set the frame ID.
⚫
Error: triggers when an error occurs to the FlexRay bus.
Press
Error
to select the error type. It includes Head CRC Err, Tail CRC Err,
Decode Err, and Random Err.
Note:
As the occurrence possibility of specified FlaxRay frame is very low, it is
recommended that you set the oscilloscope to "Normal" trigger mode when
the trigger condition is set to "Frame", so as to prevent the instrument from
triggering automatically while waiting for the specified frame. The same goes
for "Error" trigger condition. In addition, when multiple FlexRay errors occur
at the same time, you need to adjust the trigger holdoff so as to view the
specific error.
Trigger Mode:
In the trigger control area (Trigger) on the front panel, press
Mode
to quickly switch
the current trigger mode. For details, refer to descriptions in
.
Содержание MSO8000 Series
Страница 1: ...User Guide MSO8000 Series Digital Oscilloscope Mar 2021 RIGOL TECHNOLOGIES CO LTD...
Страница 2: ......
Страница 24: ......
Страница 72: ......
Страница 78: ......
Страница 148: ......
Страница 202: ......
Страница 208: ......
Страница 214: ......
Страница 248: ...RIGOL Chapter 12 Protocol Decoding 12 14 MSO8000 User Guide...
Страница 264: ...RIGOL Chapter 12 Protocol Decoding 12 30 MSO8000 User Guide Sync Error...
Страница 274: ...RIGOL Chapter 12 Protocol Decoding 12 40 MSO8000 User Guide TSS Sync Frame Frame ID PL HCRC CYC Data TCRC...
Страница 284: ......
Страница 290: ......
Страница 296: ......
Страница 302: ......
Страница 308: ......
Страница 360: ......
Страница 383: ...Chapter 21 Remote Control RIGOL MSO8000 User Guide 21 3 5 6 3 4...
Страница 394: ......