![Rigol MSO8000 Series Скачать руководство пользователя страница 178](http://html1.mh-extra.com/html/rigol/mso8000-series/mso8000-series_user-manual_1461151178.webp)
RIGOL
Chapter 6 Operations and Measurements
6-30
MSO8000 User Guide
➢
Press
Peak to Peak
to enable or disable the peak-peak priority setting.
This function is intended for the shifted signal. If there is a large deviation,
you can view the signal waveform in priority when you enable the function.
➢
Press
CH
to select "All" or "Open".
If you select "All", press
AUTO
and the system will test the 4 analog
channels (CH1-CH4) in sequence. If no signal is found on the channel, then
the channel is disabled; if a signal is found on the channel, adjust the
channel to an optimal scale to display the signal.
If you select "Open", press
AUTO
and the system will only test the enabled
channels.
➢
Press
Overlay
to enable or disable the waveform overlay display function.
If enabled, waveforms of different channels will be displayed in the same
position of the screen; if disabled, waveforms of different channels will be
displayed on the screen from top to bottom in sequence.
➢
Press
Coupling
to enable or disable the coupling hold function. If enabled,
after you press
AUTO
, the settings for the channel coupling remain
unchanged; if disabled, then the channel coupling is, by default, set to
"DC".
➢
Press
Password
, and then the password reset dialog box is displayed, as
shown in the figure below. You can reset the password. The initial password
is empty. You do not need to input the old password when you reset the
password for the first time.
➢
: Press this key to cancel the auto setting and recovers to the parameter
settings prior to pressing the
AUTO
key.
Note:
The waveform auto setting function requires that the frequency of the signal
should be greater than or equal to 35 Hz, the amplitude greater than or equal
Tip
If you lock the Auto function and then exit the auto setting function
menu, you can press
Utility
→
More
→
Auto Config
to enter the auto
setting function menu again to unlock the Auto function.
Содержание MSO8000 Series
Страница 1: ...User Guide MSO8000 Series Digital Oscilloscope Mar 2021 RIGOL TECHNOLOGIES CO LTD...
Страница 2: ......
Страница 24: ......
Страница 72: ......
Страница 78: ......
Страница 148: ......
Страница 202: ......
Страница 208: ......
Страница 214: ......
Страница 248: ...RIGOL Chapter 12 Protocol Decoding 12 14 MSO8000 User Guide...
Страница 264: ...RIGOL Chapter 12 Protocol Decoding 12 30 MSO8000 User Guide Sync Error...
Страница 274: ...RIGOL Chapter 12 Protocol Decoding 12 40 MSO8000 User Guide TSS Sync Frame Frame ID PL HCRC CYC Data TCRC...
Страница 284: ......
Страница 290: ......
Страница 296: ......
Страница 302: ......
Страница 308: ......
Страница 360: ......
Страница 383: ...Chapter 21 Remote Control RIGOL MSO8000 User Guide 21 3 5 6 3 4...
Страница 394: ......