RIGOL
8-2
MSO8000 User Guide
Power Quality
By analyzing the power quality, you can test the quality of AC input lines. The specific
measurement parameters for power quality analysis include V_RMS, I_RMS, real
power, apparent power, reactive power, power factor, reference frequency, phase
angle, impedance, voltage crest factor, and current crest factor.
Press
Measure
→
Analyze
→
Power Analyzer
to open the power analysis setting
menu, or enable the touch screen and then tap the function navigation icon
at
the lower-left corner of the screen to open the function navigation. Then, tap the
"Power" icon to open the power analysis setting menu. You can also press
Measure
→
Add
, then in the displayed parameter measurement category area, rotate the
multifunction knob
to select the "Analyze" tab, then select "UPA" to open the
power analysis setting menu.
⚫
Select the Analysis Type
In the power analysis setting menu, press
Analysis Type
and rotate the
multifunction knob
to select "Power Quality", and then press down the knob
to select it. You can also press
Analysis Type
continuously or enable the touch
screen to select the desired analysis type.
⚫
Set the Power Quality Analysis Source
In the power analysis setting menu, press
Source
to enter the source setting
menu.
➢
Press
Voltage
and then rotate the multifunction knob
to select the
channel (CH1-CH4) for acquiring voltage. Press down the knob to select it.
You can also press
Voltage
continuously or enable the touch screen to
select the channel.
➢
Press
Current
and then rotate the multifunction knob
to select the
channel (CH1-CH4) for acquiring current. Press down the knob to select it.
You can also press
Current
continuously or enable the touch screen to
select the channel.
➢
Press
Freq. Ref
continuously to select "Voltage" or "Current" as the
frequency reference.
⚫
Set the Reference Level
In the power analysis setting menu, press
Ref Levels
(level of the frequency
reference channel) to enter the reference level setting menu.
➢
Press
Type
, and rotate the multifunction knob
to set the reference
level display type to "Percent(%)" or "Absolute", and then press down the
knob to select it. You can also press
Type
continuously or enable the touch
screen to select the type.
➢
Press
Upper
, then rotate the multifunction knob
or use the pop-up
Содержание MSO8000 Series
Страница 1: ...User Guide MSO8000 Series Digital Oscilloscope Mar 2021 RIGOL TECHNOLOGIES CO LTD...
Страница 2: ......
Страница 24: ......
Страница 72: ......
Страница 78: ......
Страница 148: ......
Страница 202: ......
Страница 208: ......
Страница 214: ......
Страница 248: ...RIGOL Chapter 12 Protocol Decoding 12 14 MSO8000 User Guide...
Страница 264: ...RIGOL Chapter 12 Protocol Decoding 12 30 MSO8000 User Guide Sync Error...
Страница 274: ...RIGOL Chapter 12 Protocol Decoding 12 40 MSO8000 User Guide TSS Sync Frame Frame ID PL HCRC CYC Data TCRC...
Страница 284: ......
Страница 290: ......
Страница 296: ......
Страница 302: ......
Страница 308: ......
Страница 360: ......
Страница 383: ...Chapter 21 Remote Control RIGOL MSO8000 User Guide 21 3 5 6 3 4...
Страница 394: ......