Chapter 5 To Trigger the Oscilloscope
RIGOL
MSO8000 User Guide
5-53
descriptions in
. The current trigger source is displayed at the
upper-right corner of the screen.
Note:
Only when we select the channel (that has been input with signals) as the
trigger source, can we obtain a stable trigger.
Trigger Condition:
Press
When
, and rotate the multifunction knob
to select the desired trigger
condition, then press down the knob to select it. You can also press
When
continuously or enable the touch screen to select it.
⚫
Sync: triggers on the specified sync type. After the trigger condition is selected,
press
Sync
to select the desired sync type: Data Sync, Cmd/Status Sync, or All
Sync.
⚫
Data: triggers on the specified data word. After the trigger condition is selected,
press
Comp
to select the comparison conditions. The available choices include
=, ≠, >, <, ><, <>.
➢
=: triggers when the channel's data word equals the set data word. Press
Min
to set the lower limit of the data word. For setting methods, refer to
descriptions in
.
➢
≠: triggers when the channel's data word does not equal the set data word.
Press
Min
to set the lower limit of the data word. For setting methods, refer
to descriptions in
.
➢
<: triggers when the channel's data word is smaller than the set data word.
Press
Max
to set the upper limit of the data word. For setting methods,
refer to descriptions in
.
➢
>: triggers when the channel's data word is greater than the set data word.
Press
Min
to set the lower limit of the data word. For setting methods, refer
to descriptions in
.
➢
><: triggers when the channel's data word is smaller than the upper limit of
the data word and greater than the lower limit of the data word. Press
Max
and
Min
to set the upper limit and lower limit of the data word respectively.
For setting methods, refer to descriptions in
.
➢
<>: triggers when the channel's data word is greater than the upper limit of
the data word and smaller than the lower limit of the data word. Press
Max
and
Min
to set the upper limit and lower limit of the data word respectively.
For setting methods, refer to descriptions in
.
⚫
RTA: triggers on the specified remote terminal address. After this trigger
condition is selected, press
RTA
to set the remote terminal address. For setting
methods, refer to descriptions in
.
⚫
RTA+11Bit: triggers on the RTA and the remaining 11 bits. After this trigger
condition is selected:
➢
Press
RTA
to set the remote terminal address. For setting methods, refer to
descriptions in
.
➢
Press
Bit time
to set the bit time position value to 0 (low), 1 (high), or X
(don't care). For setting methods, refer to descriptions in
.
⚫
Error: triggers on the specified error type. After this trigger condition is selected,
Содержание MSO8000 Series
Страница 1: ...User Guide MSO8000 Series Digital Oscilloscope Mar 2021 RIGOL TECHNOLOGIES CO LTD...
Страница 2: ......
Страница 24: ......
Страница 72: ......
Страница 78: ......
Страница 148: ......
Страница 202: ......
Страница 208: ......
Страница 214: ......
Страница 248: ...RIGOL Chapter 12 Protocol Decoding 12 14 MSO8000 User Guide...
Страница 264: ...RIGOL Chapter 12 Protocol Decoding 12 30 MSO8000 User Guide Sync Error...
Страница 274: ...RIGOL Chapter 12 Protocol Decoding 12 40 MSO8000 User Guide TSS Sync Frame Frame ID PL HCRC CYC Data TCRC...
Страница 284: ......
Страница 290: ......
Страница 296: ......
Страница 302: ......
Страница 308: ......
Страница 360: ......
Страница 383: ...Chapter 21 Remote Control RIGOL MSO8000 User Guide 21 3 5 6 3 4...
Страница 394: ......