Chapter 6 Operations and Measurements
RIGOL
MSO5000 User's Guide
6-15
Press
More
Sensitivity
to set the sensitivity of the digital signal converted
from the analog signal on the source. For setting methods, refer to descriptions
in
.
Press
More
Label
to set the label for math operation results. For details,
refer to descriptions in
.
Intg
Calculates the integral of the selected source. For example, you can use integral to
measure the area under a waveform or the pulse energy.
Press
Math
Math1
Operator
to select "Intg":
Press
Operation
continuously to enable or disable the "Intg" operation
function.
Press
SourceA
, and rotate the multifunction knob
to select the channel for
SourceA. Then press down the knob to select the channel. You can also press
SourceA
continuously or enable the touch screen to select the channel. The
available channels include CH1-CH4 and Ref1-Ref10.
Note:
The available channels for Math2 are CH1-CH4, Ref1-Ref10 or Math1.
The available channels for Math3 are CH1-CH4, Ref1-Ref10, Math1 or Math2.
The available channels for Math4 are CH1-CH4, Ref1-Ref10, Math1, Math2 or
Math3. Only the channels that have been enabled currently can be selected.
Press
Offset
to set the vertical offset of the operation results. For setting
methods, refer to the descriptions in
. You can
also enable the touch screen and use the "Drag" gesture to adjust the vertical
offset.
Press
Scale
to set the vertical scale of the operation results. For setting methods,
refer to the descriptions in
. You can also enable
the touch screen and use the "Pinch & Stretch" gesture to adjust the vertical
scale.
Press
AutoScale
to adjust the vertical scale and offset of the operation results
to an optimal value based on the current configuration, so as to better observe.
Press
More
Bias
to set the DC offset correction factor for the input signal.
Note that this parameter can only be set by using the numeric keypad.
Press
More
Invert
to enable or disable the inverted display function of the
waveform.
Press
More
Label
to set the label for math operation results. For details,
refer to descriptions in
.
Press
More
Color Grade
to enable or disable the color grade display of the
operation results. When enabled, different colors are displayed on the screen to
indicate the times of data acquisition or acquisition probability.
Press
More
Reset Color Grade
to clear the color grade display and display
the color grade again.
www.
GlobalTestSupply
.com
Find Quality Products Online at:
Содержание MSO5000 Series
Страница 2: ...www GlobalTestSupply com nd Quality Products Online at sales GlobalTestSupply co...
Страница 52: ...www GlobalTestSupply com nd Quality Products Online at sales GlobalTestSupply co...
Страница 64: ...www GlobalTestSupply com nd Quality Products Online at sales GlobalTestSupply co...
Страница 70: ...www GlobalTestSupply com nd Quality Products Online at sales GlobalTestSupply co...
Страница 82: ...www GlobalTestSupply com nd Quality Products Online at sales GlobalTestSupply co...
Страница 184: ...www GlobalTestSupply com nd Quality Products Online at sales GlobalTestSupply co...
Страница 190: ...www GlobalTestSupply com nd Quality Products Online at sales GlobalTestSupply co...
Страница 196: ...www GlobalTestSupply com nd Quality Products Online at sales GlobalTestSupply co...
Страница 256: ...www GlobalTestSupply com nd Quality Products Online at sales GlobalTestSupply co...
Страница 262: ...www GlobalTestSupply com nd Quality Products Online at sales GlobalTestSupply co...
Страница 268: ...www GlobalTestSupply com nd Quality Products Online at sales GlobalTestSupply co...
Страница 274: ...www GlobalTestSupply com nd Quality Products Online at sales GlobalTestSupply co...
Страница 280: ...www GlobalTestSupply com nd Quality Products Online at sales GlobalTestSupply co...
Страница 356: ...www GlobalTestSupply com nd Quality Products Online at sales GlobalTestSupply co...