Rigol MSO5000 Series Скачать руководство пользователя страница 117

Chapter 5 To Trigger the Oscilloscope

 

 

RIGOL

 

MSO5000 User's Guide

 

5-35 

high level. 

 

Stop: triggers when SDA data transitions from low level to high level while SCL is 

high level.   

 

Restart: triggers when another start condition occurs before a stop condition. 

 

MissedAck: triggers when the SDA data is high level during any acknowledgment 

of SCL clock position.

 

 

Address: the trigger searches for the specified address value. When this event 

occurs,  the  oscilloscope  will  trigger  on  the  read/write  bit.  After  this  trigger 

condition is selected: 

 

Press 

AddrBits

 and rotate the multifunction knob 

  to select the desired 

address  bits.  Then press down  the  knob  to  select it. You can also  press 

AddrBits

 continuously or enable the touch screen to select it. The available 

address bits are "7 Bits", "8 Bits", and "10 Bits". 

 

Press 

Address

, then rotate the multifunction knob 

  or use the numeric 

keypad to set the address of I2C trigger. 

 

Press 

More 

 

Direction

 to select "Write", "Read", or "R/W".   

Note:

 This setting is not available when 

AddrBits

 is set to "8 Bits". 

 

Data: the trigger searches for the specified data value on the data line (SDA). 

When  this  event  occurs,  the  oscilloscope  will  trigger  on  the  clock  line  (SCL) 

transition edge of the last bit of data. After this trigger condition is selected: 

 

Press 

Bit X

, and set the current bit with the following two methods. 

Method 1: First rotate the multifunction knob 

  to switch to the data bit 

that needs to be operated on. Then press down the knob continuously to 

set the current bit. As shown in the figure below, the left is its binary format, 

and the right is its corresponding Hex format. 

 

 
Method 2: Press 

Bit X

 or enable the touch screen to select the data bit. 

Then set it with the pop-up virtual keypad. For setting methods, refer to 

descriptions  in 

"Pattern  Trigger"

.  If  you  modify  the  binary  data  bit,  a 

virtual keypad as shown in Figure 5-15 (a) is displayed. If you modify the 

hex data bit, a virtual keypad as shown in Figure 5-15 (b) is displayed. 

             

 

(a)

 

Virtual Keypad for Binary Data Bit    (b) Virtual Keypad for Hex Data Bit 

Figure 5-15 Virtual Keypad for Data Bit Settings 

 

www.

GlobalTestSupply

.com

Find Quality Products Online at:

[email protected]

Содержание MSO5000 Series

Страница 1: ...RIGOL User s Guide MSO5000 Series Digital Oscilloscope Oct 2018 RIGOL SUZHOU TECHNOLOGIES INC www GlobalTestSupply com nd Quality Products Online at sales GlobalTestSupply co...

Страница 2: ...www GlobalTestSupply com nd Quality Products Online at sales GlobalTestSupply co...

Страница 3: ...l previously released materials Information in this publication is subject to change without notice RIGOL shall not be liable for either incidental or consequential losses in connection with the furni...

Страница 4: ...ground lead must be connected to earth ground Do not connect the ground lead to high voltage Improper way of connection could result in dangerous voltages being present on the connectors controls or o...

Страница 5: ...t Do Not Operate in an Explosive Atmosphere To avoid personal injuries or damage to the instrument never operate the instrument in an explosive atmosphere Keep Product Surfaces Clean and Dry To avoid...

Страница 6: ...roduct DANGER It calls attention to an operation if not correctly performed could result in injury or hazard immediately WARNING It calls attention to an operation if not correctly performed could res...

Страница 7: ...ements on distribution boards circuit breakers wiring including cables bus bars junction boxes switches and socket outlets in the fixed installation and equipment for industrial use and some other equ...

Страница 8: ...rvoltage such as that caused by a bolt of lightning can reach the product Otherwise the operator might be exposed to the danger of an electric shock Installation Overvoltage Category Definitions Insta...

Страница 9: ...rounded Product Care and Cleaning Care Do not store or leave the instrument where it may be exposed to direct sunlight for long periods of time Cleaning Clean the instrument regularly according to its...

Страница 10: ...oid the release of such substances into the environment and avoid harm to human health we recommend you to recycle this product appropriately to ensure that most materials are reused or recycled prope...

Страница 11: ...radable standard configuration of 16 digital channels probe option is required to be ordered Up to 8 GSa s real time sample rate Up to 200 Mpts memory depth option High waveform capture rate over 500...

Страница 12: ...Measurements Introduces how to make math operation auto measurement and cursor measurement Chapter 7 Digital Voltmeter DVM and Frequency Counter Introduces how to use the digital voltmeter and the fr...

Страница 13: ...Function Arbitrary Waveform Generator option Chapter 18 Store and Load Introduces how to store and load the measurement results and the setting of the oscilloscope Chapter 19 System Utility Function...

Страница 14: ...example TRIG OUT 5 Knob Label Knob Label Knob Horizontal SCALE Horizontal Time Base Knob Vertical SCALE Vertical Scale Knob Horizontal POSITION Horizontal Position Knob Vertical OFFSET Vertical Offse...

Страница 15: ...Product Quick Guide User s Guide Programming Guide Data sheet etc For the latest version of this manual download it from the official website of RIGOL www rigol com www GlobalTestSupply com nd Qualit...

Страница 16: ...the Supporting Legs 1 4 To Connect to AC Power 1 4 Turn on Checkout 1 5 To Connect the Probe 1 5 Function Inspection 1 7 Probe Compensation 1 8 Front Panel Overview 1 9 Rear Panel Overview 1 10 Front...

Страница 17: ...mpedance 2 7 Waveform Invert 2 7 Amplitude Unit 2 8 Channel Delay 2 8 Offset Cal 2 9 Channel Label 2 10 Chapter 3 To Set the Horizontal System 3 1 To Adjust the Horizontal Time Base 3 2 To Adjust the...

Страница 18: ...I2C Trigger Option 5 34 SPI Trigger Option 5 36 CAN Trigger Option 5 38 FlexRay Trigger Option 5 41 LIN Trigger Option 5 43 I2S Trigger Option 5 45 MIL STD 1553 Trigger Option 5 48 Zone Trigger 5 50...

Страница 19: ...2 To Select Measurement Mode 7 3 To Set the Limits 7 3 Frequency Counter 7 4 To Enable or Disable the Frequency Counter 7 4 To Select the Measurement Source 7 4 To Select the Measurement Item 7 4 To S...

Страница 20: ...just the Ref Waveform Display 12 2 To Save to Internal Memory 12 3 To Clear the Display of the Reference Waveform 12 3 To View Details of the Reference Waveform 12 3 To Reset the Reference Waveform 12...

Страница 21: ...tput Basic Waveforms 17 2 To Output Sine 17 2 To Output Square 17 3 To Output Ramp 17 4 To Output Pulse 17 4 To Output DC 17 5 To Output Noise 17 5 Sinc 17 6 Exp Rise 17 6 Exp Fall 17 7 ECG 17 7 Gauss...

Страница 22: ...Set the GPIB Address 19 6 USB Connection 19 6 System related 19 7 Beeper 19 7 Language 19 7 System Information 19 7 Power On 19 7 Aux Output 19 8 Help 19 8 Self calibration 19 9 Auto Config 19 9 Print...

Страница 23: ...se the built in help system Contents in this chapter General Inspection Appearance and Dimensions To Prepare for Use Front Panel Overview Rear Panel Overview Front Panel Function Overview User Interfa...

Страница 24: ...he instrument resulting from shipment RIGOL would not be responsible for free maintenance rework or replacement of the instrument 2 Inspect the instrument In case of any mechanical damage missing part...

Страница 25: ...uick Start RIGOL MSO5000 User s Guide 1 3 Appearance and Dimensions Figure 1 1 Front View Unit mm Figure 1 2 Vertical View Unit mm www GlobalTestSupply com nd Quality Products Online at sales GlobalTe...

Страница 26: ...e or shipment when the instrument is not in use To fold the supporting legs To unfold the supporting legs Figure 1 3 To Adjust the Supporting Legs To Connect to AC Power The input AC power requirement...

Страница 27: ...SO5000 Series Datasheet For detailed technical information of the probes please refer to the specified Probe User s Guide Connect the passive probe 1 Connect the BNC terminal of the probe to an analog...

Страница 28: ...MSO5000 series supplies PLA2216 active logic probe option To apply to different application scenarios PLA2216 provides two connection methods to connect the signal under test For details refer to PLA...

Страница 29: ...gnal Output Terminal of the probe as shown in Figure 1 7 Figure 1 7 To Use the Compensation Signal 4 Set the probe attenuation to 10X and then press AUTO 5 Observe the waveform on the display In norma...

Страница 30: ...Probe Compensation 3 Use a nonmetallic screwdriver to adjust the low frequency compensation adjustment hole on the probe until the waveform is displayed as Perfectly compensated in the figure above W...

Страница 31: ...ommon Operation Keys 17 Probe Compensation Signal Output Terminal Ground Terminal 7 CLEAR Key 18 Analog Channel Input Terminals 8 Auto Waveform Display Key 19 Digital Channel Input Interface 2 9 RUN S...

Страница 32: ...send the SCPI commands the user defined programming or Web Control to control the instrument 4 LAN Interface Connect the instrument to network via this interface The instrument is in compliance with t...

Страница 33: ...AC power source 7 Kensington Lock Hole You can lock the instrument to a fixed location by using the security lock please purchase it by yourself via the lock hole Tip After the oscilloscope is connec...

Страница 34: ...th the reference waveform to locate the circuit failure LA indicates the logic analyzer key Press this key to open the logic analyzer control menu You can enable or disable any channel or channel grou...

Страница 35: ...ing status label will change accordingly Press down this knob to quickly switch the vertical scale adjustment mode between Coarse and Fine Horizontal Zoom indicates the delayed sweep key Press this ke...

Страница 36: ...oscope provides various trigger types Mode Press this key to switch the trigger mode to Auto Normal or Single Force Press this key to generate a trigger signal forcibly Trigger LEVEL modifies the trig...

Страница 37: ...ote that XY cursor mode is only available when the horizontal time base is set to XY Display Press this key to enter the display setting menu You can set the display type persistence time wave intensi...

Страница 38: ...Clear Press this key to clear all the waveforms on the screen If the oscilloscope is in the RUN state new waveforms will continue being displayed Auto Press this key to enable the waveform auto setti...

Страница 39: ...able screen brightness ranges from 1 to 100 Turn it clockwise to increase the brightness and turn it counterclockwise to decrease the brightness Press Display Intensity to adjust the waveform brightne...

Страница 40: ...e The color of the currently selected waveform of the digital channel is consistent with that of the channel label being displayed in red The grouping setting function under the logic analyzer functio...

Страница 41: ...position of the screen waveform and that of the waveform in the memory 8 Run Stop Label In the run state the label is displayed in green while in the stop state it is displayed in red You can enable...

Страница 42: ...icon the USB storage device icon time and the LAN connection icon Beeper icon Press Utility Beeper to turn on or off the beeper When the beeper is turned on will be displayed when off will be display...

Страница 43: ...annel coupling is displayed if you enable the bandwidth limit is displayed 21 CH2 Status Label Displays the status of CH2 Displays the vertical scale of CH2 That is the voltage value per grid of CH2 i...

Страница 44: ...nable the function navigation Tap the displayed numeric keypad to set the parameters Tap the virtual keypad to set the label name and the filename Tap the close button at the upper right corner of the...

Страница 45: ...or offset Drag the window controls to change the positions of the window e g numeric keypad Drag the marker to change the position of the marker Figure 1 15 Drag Gesture Rectangle Drawing Enable the f...

Страница 46: ...ontal direction Select Vertical zoom expands the waveforms in the vertical direction Select Waveform zoom expands the waveforms both in the horizontal and vertical direction Tip Tap the Draw rect icon...

Страница 47: ...n the numeric keypad rotate the multifunction knob to select the parameter then press down the knob to select it Then input the value or the unit You can also enable the touch screen and tap the value...

Страница 48: ...yourself as shown in Figure 1 18 The method is as follows align the lock with the lock hole and plug it into the lock hole vertically turn the key clockwise to lock the oscilloscope and then pull the...

Страница 49: ...e help interface mainly consists of two sections The left section lists Help Options and the right section is the Help Display Area Figure 1 19 Help Information After opening the help interface you ca...

Страница 50: ...om the option list 1 Press Utility System Help and then the Help menu is displayed on the screen You can also use the touch screen function tap the function navigation icon at the lower left corner of...

Страница 51: ...tion menu and then press About to obtain the serial number and the verification code Click Generate to obtain the download link of the option license file If you need to use the option license file pl...

Страница 52: ...www GlobalTestSupply com nd Quality Products Online at sales GlobalTestSupply co...

Страница 53: ...the vertical systems of the four channels are the same This chapter takes CH1 as an example to introduce the setting method for the vertical system For this oscilloscope only CH1 is enabled by defaul...

Страница 54: ...ot activated the channel status label is shown in the following figure To activate the channel press 1 on the front panel in the vertical control area Vertical or enable the touch screen to tap the wa...

Страница 55: ...retch gesture on the touch screen For details refer to the Pinch Stretch section Enable the touch screen function and tap the channel status label at the bottom of the screen The following window is d...

Страница 56: ...ms of the corresponding channel moves up and down The vertical offset information as shown in the following figure in the channel status label at the bottom of the screen will change accordingly The a...

Страница 57: ...t coupling mode is displayed in the channel status label at the bottom of the screen as shown in the figure below You can also enable the touch screen and then touch the CH1 setting menu to select the...

Страница 58: ...M MSO5354 20 M 100 M 200 M Note Bandwidth limit can not only reduce the noise but also can attenuate or eliminate the high frequency components of the signal Probe Ratio MSO5000 series allows you to...

Страница 59: ...e menu is grayed out and you are not allowed to set it manually Waveform Invert Press 1 to open the setting menu of CH1 Then press Invert continuously to enable or disable the waveform invert When ena...

Страница 60: ...e Unit press Unit continuously or rotate the multifunction knob to select a unit You can also enable the touch screen to select the desired unit with touch gestures The default unit for the value is V...

Страница 61: ...ponent or external environment disturbance may occur on the zero cross voltage of the channel which will affect the measurement results of the vertical parameters MSO5000 allows you to set an offset c...

Страница 62: ...s Label and the label editing interface is automatically displayed as shown in Figure 2 3 You can input the label manually For the label input method refer to To Create a Folder Figure 2 3 Label Editi...

Страница 63: ...acters input the desired characters again Enable the touch screen use the touch gestures to directly move the cursor to the right side of the character required to be deleted or modified and then tap...

Страница 64: ...www GlobalTestSupply com nd Quality Products Online at sales GlobalTestSupply co...

Страница 65: ...000 User s Guide 3 1 Chapter 3 To Set the Horizontal System Contents in this chapter To Adjust the Horizontal Time Base To Adjust the Horizontal Position Delayed Sweep www GlobalTestSupply com nd Qual...

Страница 66: ...reen For details refer to the Pinch Stretch section Enable the touch screen function and tap the horizontal time base label at the top of the screen The following window is displayed Tap the icon at t...

Страница 67: ...on at the right upper corner of the screen changes accordingly as shown in the figure below You can adjust the horizontal position with the following three methods Rotate Horizontal POSITION to adjust...

Страница 68: ...etails In the horizontal control area on the front panel press Zoom to enable or disable the delayed sweep By default it is disabled You can also press Acquire More Zoom to enable or disable the delay...

Страница 69: ...ate the Horizontal SCALE knob to set the required main time base and then enable the delayed sweep function Waveform after enlargement The waveform in the lower part of the screen is the horizontally...

Страница 70: ...www GlobalTestSupply com nd Quality Products Online at sales GlobalTestSupply co...

Страница 71: ...er 4 To Set the Sample System Contents in this chapter Timebase Mode Acquisition Mode Sampling Mode Sample Rate LA Sample Rate Memory Depth LA Memory Depth Anti Aliasing Horizontal Expansion www Globa...

Страница 72: ...nly when this mode is enabled can delayed sweep be turned on In this mode when the horizontal time base is equal to or greater than 200 ms div the instrument enters slow sweep mode For details refer t...

Страница 73: ...t and output signals of the circuit Application example measures the phase deviation between the input signals of two channels Use Lissajous method 1 Connect one sine signal to CH1 and then connect an...

Страница 74: ...Auto ROLL to select ON The system automatically enters the ROLL mode and there is no slow sweep Note If the current delayed sweep is enabled then when you enable the ROLL mode the delayed sweep is dis...

Страница 75: ...eforms from multiple samples to reduce the random noise of the input signal and improve the vertical resolution Greater number of averages can lower the noise and increase the vertical resolution whil...

Страница 76: ...s the maximum and minimum values of the signal within the sample interval to get the envelope of the signal or the narrow pulse that might be lost In this mode signal aliasing can be prevented but the...

Страница 77: ...is 8 GSa s The maximum real time sample rate in the half channel mode is 4 GSa s and the maximum real time sample rate in the all channel mode is 2 GSa s Note The sample rate is displayed both in the...

Страница 78: ...maller than the actual signal frequency 3 Waveform Leakage when the sample rate is too low the waveform rebuilt from the sample data does not reflect all the actual signal information LA Sample Rate L...

Страница 79: ...e rate rotate the Horizontal SCALE knob to adjust the horizontal time base or modify the memory depth Memory Depth Memory depth refers to the number of points of the oscilloscope that can store in one...

Страница 80: ...omatically according to the current sample rate LA Memory Depth Press Acquire More to view the memory depth of the current digital channel under LA Mem Depth The LA memory depth will change with the m...

Страница 81: ...en the horizontal time base is modified the waveform will be expanded or compressed horizontally relative to the rightmost position of the screen Trigger when the horizontal time base is modified the...

Страница 82: ...www GlobalTestSupply com nd Quality Products Online at sales GlobalTestSupply co...

Страница 83: ...very time base sweep or acquisition starts from the user defined trigger condition namely every sweep is synchronous with the acquisition and the waveforms acquired is overlapped so as to display the...

Страница 84: ...ignals input from analog channels CH1 CH4 can all be used as trigger sources No matter whether the channel selected is enabled the channel can work normally Digital channel input Signals input from di...

Страница 85: ...on is displayed at the lower left side of the screen as shown in the figure below H indicates the upper limit of the trigger level L indicates the lower limit of the trigger level and indicates the de...

Страница 86: ...ta acquired just before the trigger Then the oscilloscope will fill the post trigger buffer and display the data in the acquisition memory If the acquisition is activated via RUN STOP the oscilloscope...

Страница 87: ...ent waveform data will not overwrite the current waveform After a single trigger mode is initiated the operating status of the oscilloscope is in STOP state Note In Normal and Single trigger modes pre...

Страница 88: ...shown in the figure below the holdoff time should be set to a value that is greater than t1 and smaller than t2 Figure 5 2 Schematic Diagram of Trigger Holdoff In the trigger control area Trigger on t...

Страница 89: ...More Noise Reject to enable or disable noise rejection Note This function is only available when the trigger source is an analog channel Trigger Type MSO5000 series oscilloscope provides the following...

Страница 90: ...able trigger Edge Type Press Slope and then rotate the multifunction knob to select the input signal the edge on which the oscilloscope triggers Then the current edge type is displayed at the upper ri...

Страница 91: ...as shown in the figure below Figure 5 3 Positive Pulse Width Negative Pulse Width Trigger Type Press Type and then rotate the multifunction knob to select Pulse Press down the knob to select the trig...

Страница 92: ...ers when the negative pulse width of the input signal is greater than the specified pulse width When you select Negative for polarity for trigger condition the oscilloscope triggers when the negative...

Страница 93: ...ed as the time difference between the two crossing points of trigger level line A and B with the falling edge as shown in the figure below Figure 5 4 Positive Slope Time Negative Slope Time Trigger Ty...

Страница 94: ...When you select Falling for the edge type for trigger condition the oscilloscope triggers when the negative slope time of the input signal is greater than the specified time When you select Falling f...

Страница 95: ...ent level adjustment type Rotate the Trigger LEVEL knob to adjust the corresponding trigger level During the adjustment of trigger level two trigger level lines appear on the screen and they move up a...

Страница 96: ...ource can we obtain a stable trigger Video Polarity Press Polarity continuously to select the desired video polarity The polarities available are positive polarity and negative polarity Video Standard...

Страница 97: ...field Noise Rejection Set the trigger parameter noise rejection under this trigger type For details refer to descriptions in Noise Rejection Trigger Mode In the trigger control area Trigger on the fro...

Страница 98: ...igure below You can also press Type continuously to select the trigger type or enable the touch screen to tap the desired trigger type and select it Source Selection Press Source to open the signal so...

Страница 99: ...to the left or tap the right arrow key to move the cursor to the right Then select a pattern After the pattern of the current channel is set completely the cursor goes to the next pattern bit automat...

Страница 100: ...corner of the screen Duration Trigger In duration trigger the instrument identifies a trigger condition by searching for the duration of a specified pattern This pattern is a logical AND combination o...

Страница 101: ...selected to L i g the voltage level is lower than the trigger level threshold level of the channel X sets the pattern of the channel selected to X i g this channel is not used as a part of the patter...

Страница 102: ...criptions in Trigger Mode Trigger Parameter Setting Set the trigger parameters trigger holdoff and noise rejection under this trigger type For details refer to descriptions in Trigger Holdoff and Nois...

Страница 103: ...type Rising starts timing when the rising edge of the input signal passes through the trigger level Falling starts timing when the falling edge of the input signal passes through the trigger level Eit...

Страница 104: ...ht corner of the screen Note Only when we select the channel that has been input with signals as the trigger source can we obtain a stable trigger Polarity Press Polarity continuously to select the pu...

Страница 105: ...he upper and lower limits of the trigger level at the same time and the trigger level deviation the difference between the upper limit and lower limit of the trigger level remains unchanged Note In Ru...

Страница 106: ...a stable trigger Edge Type Press Slope and then rotate the multifunction knob to select the input signal the edge on which the oscilloscope triggers You can also press the Slope key continuously or en...

Страница 107: ...the upper limit and lower limit of the trigger level remains unchanged Note In Window trigger pressing Trigger LEVEL can quickly switch the current level adjustment type Rotate the Trigger LEVEL knob...

Страница 108: ...touch screen to tap the desired trigger type and select it Source A Press SourceA to select CH1 CH4 or D0 D15 as the trigger source of Source A For details please refer to descriptions in Trigger Sou...

Страница 109: ...r limit in Delay trigger Note The lower time limit must be smaller than the upper time limit triggers when the time difference T between the specified edges of Source A and Source B is smaller than th...

Страница 110: ...Then the current trigger setting information is displayed at the upper right corner of the screen as shown in the figure below You can also press Type continuously to select the trigger type or enabl...

Страница 111: ...time is smaller than the specified hold time After selecting this type press Hold and rotate the multifunction knob to set the hold time You can also use the pop up numeric keypad to set it Setup Hold...

Страница 112: ...setting information is displayed at the upper right corner of the screen as shown in the figure below You can also press Type continuously to select the trigger type or enable the touch screen to tap...

Страница 113: ...igger level threshold level value is displayed at the upper right corner of the screen RS232 Trigger Option RS232 bus is a serial communication mode used in data transmission between PCs or between a...

Страница 114: ...f bits per frame Press More Data Bits then rotate the multifunction knob to select the desired data bits Then press down the knob to select You can also press Data Bits continuously or enable the touc...

Страница 115: ...te at a small step of 1 bps Trigger Mode In the trigger control area Trigger on the front panel press Mode to quickly switch the current trigger mode For details refer to descriptions in Trigger Mode...

Страница 116: ...formation is displayed at the upper right corner of the screen as shown in the figure below You can also press Type key continuously to select the trigger type or enable the touch screen to tap the de...

Страница 117: ...s not available when AddrBits is set to 8 Bits Data the trigger searches for the specified data value on the data line SDA When this event occurs the oscilloscope will trigger on the clock line SCL tr...

Страница 118: ...er conditions Trigger Mode In the trigger control area Trigger on the front panel press Mode to quickly switch the current trigger mode For details refer to descriptions in Trigger Mode Trigger Parame...

Страница 119: ...select the desired trigger condition Timeout the oscilloscope starts to search for the data SDA on which to trigger after the clock signal SCL stays in the idle state for a specified period of time Af...

Страница 120: ...type e g Answer Error Check Error Format Error etc of the CAN signal The data frame format of the CAN bus is as shown in the figure below Figure 5 17 Data Frame Format of the CAN Bus Trigger Type Pre...

Страница 121: ...ad to set the length of the data Its range is from 1 to 8 Data ID triggers on the data frames with the specified ID and data Set the following parameters Press Define to select Data or ID When you sel...

Страница 122: ...he start of the bit to the sample position to the bit time as shown in the figure below Figure 5 18 Sample Position Press More Sample Position then rotate the multifunction knob or use the pop up nume...

Страница 123: ...ayed at the upper right corner of the screen as shown in the figure below You can also press Type continuously to select the trigger type or enable the touch screen to tap the desired trigger type and...

Страница 124: ...in and Count Max Press More Cyc Comp to select the comparison conditions The available choices include Press More Count Max or press More Count Min then rotate the multifunction knob or use the pop up...

Страница 125: ...e Format of the LIN Bus Trigger Type Press Type and then rotate the multifunction knob to select LIN Press down the knob to select the trigger type Then the current trigger setting information is disp...

Страница 126: ...or use the pop up numeric keypad to set ID Sleep triggers when the sleep frame is found Wakeup triggers when the wakeup frame is found Error triggers on the specified type of error frame Press Error...

Страница 127: ...the trigger level threshold level Refer to Trigger LEVEL Threshold Level The current trigger level threshold level value is displayed at the upper right corner of the screen I2S Trigger Option In I2S...

Страница 128: ...t corner of the screen Note Only when we select the channel that has been input with signals as the trigger source can we obtain a stable trigger Edge Type Press SCLK Edge continuously or rotate the m...

Страница 129: ...to descriptions in I2C Trigger Option Width Press More Width then rotate the multifunction knob or use the pop up numeric keypad to set the width Its range is from 4 to 32 User Width Press More User W...

Страница 130: ...ouch screen to tap the desired trigger type and select it Source Selection Press Source to open the signal source list and select CH1 CH4 For details refer to descriptions in Trigger Source The curren...

Страница 131: ...it triggers on the RTA and the remaining 11 bits After this trigger condition is selected Press RTA to set the remote terminal address For setting methods refer to descriptions in I2C Trigger Option P...

Страница 132: ...ss More Polarity to select the desired polarity The polarities available are positive polarity and negative polarity Trigger Mode In the trigger control area Trigger on the front panel press Mode to q...

Страница 133: ...the desired trigger source and select it The available sources include CH1 CH4 The zone name color and the zone color are the same as the color of the selected source channel Set the conditions for T...

Страница 134: ...ctor each time a trigger is generated by the oscilloscope If this signal is connected to a waveform display device to measure the frequency the measurement result is the same as the current capture ra...

Страница 135: ...ons and Measurements MSO5000 can perform math operations auto measurements and cursor measurements on sampled and displayed data Contents in this chapter Math Operation Auto Measurement Cursor Measure...

Страница 136: ...creen The results of the math operation can be further measured This oscilloscope supports enabling the four math operations at the same time as shown in the figure below MSO5000 series can realize mu...

Страница 137: ...ter to observe Press More Invert to enable or disable the inverted display function of the waveform Press More Expand to expand or compress the math operation waveform around the Center or GND Press M...

Страница 138: ...compress the math operation waveform around the Center or GND Press More Label to set the label for math operation results For details refer to descriptions in Math Operation Label Press More Color Gr...

Страница 139: ...ns in Math Operation Label Press More Color Grade to enable or disable the color grade display of the operation results When enabled different colors are displayed on the screen to indicate the times...

Страница 140: ...ress More Label to set the label for math operation results For details refer to descriptions in Math Operation Label Press More Color Grade to enable or disable the color grade display of the operati...

Страница 141: ...just the vertical scale and offset of the operation results to the optimal value according to the current configuration so as to better observe Press More Unit to select the vertical unit The vertical...

Страница 142: ...hen rotate the multifunction knob or use the pop up numeric keypad to set the excursion of the peak Press Table Order then rotate the multifunction knob to select the sorting mode of the peak Press do...

Страница 143: ...h Rectangular Sine periodic and narrow band random noise Hamming A little bit better frequency resolution than Hanning Transient or short pulse the signal levels before and after the multiplication ar...

Страница 144: ...SourceB are CH1 CH4 and D0 D15 Note Only the channels that have been enabled currently can be selected If Source A or Source B selects CH1 press More Thre CH1 to set the threshold of Source A or Sourc...

Страница 145: ...digital signal converted from the analog signal on the source For setting methods refer to descriptions in Parameter Setting Method Press More Label to set the label for math operation results For de...

Страница 146: ...ction knob to adjust the vertical offset of the operation results You can also enable the touch screen and use the Drag gesture to adjust the vertical offset Press Size and then rotate the multifuncti...

Страница 147: ...n logic operation For setting methods refer to descriptions in Parameter Setting Method If Source A or Source B selects CH4 press More Thre CH4 to set the threshold of Source A or Source B in logic op...

Страница 148: ...eter Setting Method If Source A or Source B selects CH2 press More Thre CH2 to set the threshold of Source A or Source B in logic operation For setting methods refer to descriptions in Parameter Setti...

Страница 149: ...selected Press Offset to set the vertical offset of the operation results For setting methods refer to the descriptions in Parameter Setting Method You can also enable the touch screen and use the Dra...

Страница 150: ...cale to set the vertical scale of the operation results For setting methods refer to the descriptions in Parameter Setting Method You can also enable the touch screen and use the Pinch Stretch gesture...

Страница 151: ...can also enable the touch screen and use the Drag gesture to adjust the vertical offset Press Scale to set the vertical scale of the operation results For setting methods refer to the descriptions in...

Страница 152: ...so enable the touch screen and use the Drag gesture to adjust the vertical offset Press Scale to set the vertical scale of the operation results For setting methods refer to the descriptions in Parame...

Страница 153: ...utoScale to adjust the vertical scale and offset of the operation results to an optimal value based on the current configuration so as to better observe Press More Invert to enable or disable the inve...

Страница 154: ...to indicate the times of data acquisition or acquisition probability Press More Reset Color Grade to clear the color grade display and display the color grade again Abs Calculates the absolute value o...

Страница 155: ...tate the multifunction knob to select the channel for SourceA Then press down the knob to select the channel You can also press SourceA continuously or enable the touch screen to select the channel Th...

Страница 156: ...le the touch screen to select the channel The available channels for SourceA are CH1 CH4 and Ref1 Ref10 Note The available channels for Math2 are CH1 CH4 Ref1 Ref10 or Math1 The available channels for...

Страница 157: ...1 Ref10 or Math1 The available channels for Math3 are CH1 CH4 Ref1 Ref10 Math1 or Math2 The available channels for Math4 are CH1 CH4 Ref1 Ref10 Math1 Math2 or Math3 Only the channels that have been en...

Страница 158: ...channels for Math4 are CH1 CH4 Ref1 Ref10 Math1 Math2 or Math3 Only the channels that have been enabled currently can be selected Press Offset to set the vertical offset of the operation results For...

Страница 159: ...offset of the operation results For setting methods refer to the descriptions in Parameter Setting Method You can also enable the touch screen and use the Drag gesture to adjust the vertical offset Pr...

Страница 160: ...isplay of the waveform If it is enabled the label will be displayed at the left side of the waveform Press Library to select the preset waveform label including ADD SUB MUL DIV FFT AND OR XOR NOT Intg...

Страница 161: ...ttom of the screen Press this key and then one rising edge of the signal is displayed automatically on the screen Meanwhile the system will make measurements for the rise time of the currently display...

Страница 162: ...s of different channels will be displayed on the screen from top to bottom in sequence Press Coupling to enable or disable the coupling hold function If enabled after you press AUTO the settings for t...

Страница 163: ...ult is not within the valid range too large or too small then the measurement results are invalid and is displayed on the screen Please re input the signal or set the signal Time Parameters Figure 6 1...

Страница 164: ...ddle value and threshold low limit are 90 50 and 10 respectively You can press Measure Setting to enter the sub menu to set the parameters Count Values 1 Positive Pulse Count The number of positive pu...

Страница 165: ...eters Delay and Phase Source A Source B Period Delay 1 Delay A B indicates the time difference between the threshold middle values of the rising edge of Source A and that of Source B Negative delay in...

Страница 166: ...d of Source A 6 Phase A B indicates the phase deviation between the threshold middle values of the falling edge of Source A and that of Source B It is expressed in degree The formula is as follows 360...

Страница 167: ...point to the lowest point of the waveform 4 Vtop indicates the voltage value from the flat top of the waveform to the GND 5 Vbase indicates the voltage value from the flat base of the waveform to the...

Страница 168: ...i x is the amplitude of the i th point Average is the waveform average value and n is the number of the points being measured Other Parameters 1 Positive Slew Rate On the rising edge first calculate...

Страница 169: ...ower Vavg VRMS Per VRMS Overshot Preshoot Area Per Area and Variance Other measurement parameters Delay A B Delay A B Delay A B Delay A B Phase A B Phase A B Phase A B and Phase A B In the measurement...

Страница 170: ...a digital channel Source B can also be set to an analog channel or a digital channel 3 Indicator Press Indicator to enable or disable the indicator You can also press Measure Remove Indicator to enabl...

Страница 171: ...asurement For details refer to descriptions in Parameter Setting Method Increasing the lower limit to the current middle value will automatically increase the middle value and upper limit to keep them...

Страница 172: ...l to remove all the displayed measurement items Press Indicator to enable or disable the indicator Besides you can also press Measure Add Remove to enter the Remove sub menu Statistical Function Perfo...

Страница 173: ...screen Press Measure All Measure to select OFF disabling the all measurement function or select the desired channel for measurement CH1 CH4 If you select CH1 data of all measurement parameters of CH1...

Страница 174: ...s a horizontal dotted line is displayed at the right section of the screen In the X Y cursor mode Y cursor is used to measure the waveform amplitude of CH2 Press Cursor on the front panel to open the...

Страница 175: ...value at Cursor A BX indicates the X value at Cursor B BY indicates the Y value at Cursor B X indicates the horizontal spacing between Cursor A and Cursor B Y indicates the vertical spacing between C...

Страница 176: ...unction knob to adjust the vertical position of Cursor A Y cursor Press BY then rotate the multifunction knob to adjust the vertical position of Cursor B Y cursor Press AY BY then rotate the multifunc...

Страница 177: ...A and Cursor B are respectively defined as 0 and 360 Meanwhile two vertical cursors in blue that cannot be moved are displayed on the screen as the phase reference position the phase position at 0 an...

Страница 178: ...itions of Cursor A and Cursor B are respectively defined as 0 and 100 Meanwhile two cursors in blue that cannot be moved are displayed on the screen as the ratio reference position the ratio position...

Страница 179: ...esults will be changed accordingly AX indicates the X value at Cursor A AY indicates the Y value at Cursor A BX indicates the X value at Cursor B BY indicates the Y value at Cursor B X indicates the h...

Страница 180: ...sition of Y cursor Press AY then rotate the multifunction knob to adjust the vertical position of Cursor A Y cursor Press BY then rotate the multifunction knob to adjust the vertical position of Curso...

Страница 181: ...just the cursor positions to measure the X and Y values at the crossing points of the two pairs of cursors Press Mode to select XY to enable the XY cursor measurement function The measurement results...

Страница 182: ...ction knob to adjust the Y value at Cursor A and the Y value at Cursor B simultaneously During the adjustment the measurement results will change accordingly The adjustable range is limited within the...

Страница 183: ...rs are displayed on the screen Before enabling the indicator you need to enable at least one parameter for auto measurement The number of cursors will change along with the enabled measurement paramet...

Страница 184: ...www GlobalTestSupply com nd Quality Products Online at sales GlobalTestSupply co...

Страница 185: ...series oscilloscope provides a built in digital voltmeter DVM and frequency counter which enable you to perform an accurate measurement improving user experience in counter and frequency measurement...

Страница 186: ...s DVM continuously to enable or disable DVM measurement When enabled the following figure is displayed The DVM display also has a scale that is determined by the channel s vertical scale and vertical...

Страница 187: ...o enter the sub menu of Limits Then set the following parameters Beeper Press Beeper continuously to enable or disable the beeper Limits Condition Setting Press When then rotate the multifunction knob...

Страница 188: ...press Counter continuously to enable or disable the frequency counter measurement When enabled the following figure is displayed To Select the Measurement Source Press Source then rotate the multifun...

Страница 189: ...e resolution the longer the gate time In this way the measurement time will be longer To Clear Count When Totalize is selected under Measure the oscilloscope measures the count of edge events on the s...

Страница 190: ...www GlobalTestSupply com nd Quality Products Online at sales GlobalTestSupply co...

Страница 191: ...alysis function which can help you easily analyze the efficiency and reliability of the switch mode power supplies With the power analysis function you can analyze the power quality and output ripple...

Страница 192: ...el CH1 CH4 for acquiring voltage Press down the knob to select it You can also press Voltage continuously or enable the touch screen to select the channel Press Current and then rotate the multifuncti...

Страница 193: ...and execute statistics on the measurement results again Open or Close the Results Display Window Press Display continuously to enable or disable the display of the results When enabled the statistica...

Страница 194: ...r enable the touch screen to select the desired analysis type Set the Ripple Analysis Source In the power analysis setting menu press Source and rotate the multifunction knob to select the ripple anal...

Страница 195: ...ess Tips and then the connection diagram of the ripple analysis is displayed on the screen Please connect the cables according to the connection method as shown in Figure 8 2 Press Tips again or enabl...

Страница 196: ...www GlobalTestSupply com nd Quality Products Online at sales GlobalTestSupply co...

Страница 197: ...rend of waveforms and quickly locate the potential problems of the signal Contents in this chapter To Enable or Disable the Histogram Function To Select the Histogram Type To Select the Histogram Sour...

Страница 198: ...nu is displayed at the right section of the screen Refer to descriptions in Rectangle Drawing to enter the Histogram setting menu Press Enable continuously to enable or disable the histogram analysis...

Страница 199: ...t The available range of height is from 1 to 4 By default it is 2 To Set the Histogram Range When Horizontal or Vertical is selected under Type you need to set the window range Set Left Limit Right Li...

Страница 200: ...Pk_Pk indicates the Delta between the max value and the min value Mean indicates the average value of the histogram Median indicates the median value of the histogram Mode indicates the mode value of...

Страница 201: ...esign and software design Before using the digital channels connect the oscilloscope and the device under test using PLA2216 active logic probe option To apply to different application scenarios PLA22...

Страница 202: ...ptions in Group Setting To Enable Disable the Digital Channel Press On Off to enter the on off sub menu of the digital channel Enable disable the channels D7 D0 simultaneously Press D7 D0 continuously...

Страница 203: ...touch screen to tap the specified group and select it The enabled channel groups are marked with and the disabled channel groups are marked with Note This function is only valid when the channels hav...

Страница 204: ...0 the waveforms on the screen are D15 D0 in sequence from top to bottom To Set the Waveform Display Size Press Size and rotate the multifunction knob to set the waveform display size Press down the kn...

Страница 205: ...the operations for the 4 user defined groups Group1 Group2 Group3 and Group4 are the same Here take Group1 as an example Press Group1 to open the D0 D15 channel selection list the status icon is locat...

Страница 206: ...nction knob to set the color of high logic level Press down the knob to select it You can also press High continuously or enable the touch screen to select it Three colors are available and Press Edge...

Страница 207: ...channels D0 D15 As the decoding functions and setting methods of Decode1 Decode2 Decode3 and Decode4 are the same this chapter takes Decode1 as an example for illustration Press Decode on the front pa...

Страница 208: ...le the decoding function 2 Clock setting CLK Press Clock to enter the clock line setting menu Set clock channel Press CLK then rotate the multifunction knob to select any clock channel Press down the...

Страница 209: ...cannot modify them CH2 1 0 CH2 Width Bit X and CH are set automatically and you cannot modify them CH3 1 0 CH3 Width Bit X and CH are set automatically and you cannot modify them CH4 1 0 CH4 Width Bit...

Страница 210: ...Press Bit X then rotate the multifunction knob or use the pop up numeric keypad to set the bit of the channel By default 0 is selected Its available range is from 0 to width 1 Press CH then rotate the...

Страница 211: ...or disable the label display of the Parallel decoding bus When enabled the bus label Parallel will be displayed at the upper left side of the bus when the bus display is enabled Event table Press Eve...

Страница 212: ...e multifunction knob to select the form of the event table to be viewed Press down the knob to select it You can also press View continuously or enable the touch screen to select it The available choi...

Страница 213: ...e and rotate the multifunction knob to select a decode module then press down the knob to select it You can also press Decode continuously or enable the touch screen to select it 5 Noise rejection Pre...

Страница 214: ...ouch screen to select it 1 Enable or disable the bus Press Bus Status to enable or disable the decoding function 2 Quickly apply RS232 trigger settings to RS232 decoding Press Copy Trig to copy the cu...

Страница 215: ...t the polarity of data transmission It can be set to positive or negative Normal Use negative logic High level indicates logic 0 low level indicates logic 1 Invert Use positive logic High level indica...

Страница 216: ...SB Package Press Package to enable or disable the package display in data transmission When enabled several data blocks will be combined based on the separator Separator Press Separator and rotate the...

Страница 217: ...ble to close the event table Besides when the operating status is STOP you can also rotate the multifunction knob to select a specified line and view the corresponding decoding information Note When y...

Страница 218: ...played in the center of the screen Decode The setting of this menu enables you to quickly view the corresponding event tables of four decode modules Decode1 Decode2 Decode3 and Decode4 Press Decode an...

Страница 219: ...eport information its display form is related to the value of the horizontal time base is displayed in the following figure Parity Error and EOF Error When both the parity error and the EOF error are...

Страница 220: ...ion SCL and SDA For analog channels the threshold level settings will also be copied 3 Source setting Press Sources to enter the source setting menu Set the clock channel source and the threshold Pres...

Страница 221: ...Bin and ASCII Adjust the vertical position of the bus Press Position and then rotate the multifunction knob to adjust the vertical display position of the bus Set the label display Press Label to enab...

Страница 222: ...view type Press View then rotate the multifunction knob to select the form of the event table to be viewed Press down the knob to select it You can also press View continuously or enable the touch sc...

Страница 223: ...the corresponding event tables of four decode modules Decode1 Decode2 Decode3 and Decode4 Press Decode and rotate the multifunction knob to select a decode module then press down the knob to select it...

Страница 224: ...if the source is an analog channel the oscilloscope will also judge each data point logic 1 or logic 0 according to the preset threshold level Master Slave MOSI SCLK MISO CS SCLK SDA MISO MOSI Figure...

Страница 225: ...t The analog channel CH1 CH4 and digital channel D0 D15 can all be selected as the CS channel Press CS Polarity continuously to select Positive or Negative as the polarity for CS channel When you sele...

Страница 226: ...t be both set to OFF at the same time 5 Data setting Press Settings to enter the data setting menu Endian setting Press Endian to select LSB or MSB By default it is MSB LSB indicates Least Significant...

Страница 227: ...ata MISO and or MOSI the corresponding line number time and error information MISO and or MOSI Note When the source channel of MISO or MOSI is set to OFF information of the data line is not displayed...

Страница 228: ...le If you select Details the detailed data in the specified row will be displayed in the event table If appears in the row of data it indicates that the decoded data are not fully displayed At this ti...

Страница 229: ...select the desired source channel and then press down the knob to select it You can also press Source continuously or enable the touch screen to select it The available channels include CH1 CH4 and D...

Страница 230: ...t the vertical display position of the bus Set the label display Press Label to enable or disable the label display of the LIN decoding bus When enabled the bus label LIN will be displayed at the uppe...

Страница 231: ...multifunction knob to select the form of the event table to be viewed Press down the knob to select it You can also press View continuously or enable the touch screen to select it The available choic...

Страница 232: ...nction knob to select a decode module then press down the knob to select it You can also press Decode continuously or enable the touch screen to select it 6 Interpret the decoded LIN data Break Sync B...

Страница 233: ...checksum error is detected CRC is displayed in red as shown in the figure below Sync Error If the sync error is detected SYNC is displayed in red as shown in the figure below Parity Error Checksum Err...

Страница 234: ...uickly apply CAN trigger settings to CAN decoding Press Copy Trig to copy the CAN trigger settings and apply them to CAN decoding function set the corresponding CAN decoding parameters automatically F...

Страница 235: ...The CAN differential bus signals connected to an analog source channel by using a differential probe Connect the probe s positive lead to the CAN_H bus signal and connect the negative lead to the CAN...

Страница 236: ...ed the bus label CAN will be displayed at the upper left side of the bus when the bus display is enabled Event table Press Event Table to enter the event table setting menu The event table displays th...

Страница 237: ...multifunction knob to select the form of the event table to be viewed Press down the knob to select it You can also press View continuously or enable the touch screen to select it The available choice...

Страница 238: ...of four decode modules Decode1 Decode2 Decode3 and Decode4 Press Decode and rotate the multifunction knob to select a decode module then press down the knob to select it You can also press Decode cont...

Страница 239: ...lso be copied 3 Set the source and the threshold Press Source and rotate the multifunction knob to select the desired source channel and then press down the knob to select it You can also press Source...

Страница 240: ...ob or use the pop up numeric keypad to set the sample position The settable range is from 10 to 90 5 Display related setting Press Display to enter the display setting menu Set the display format Pres...

Страница 241: ...ate the multifunction knob to select a specified line and view the corresponding decoding information Figure 11 13 FlexRay Decoding Event Table Note When you adjust the horizontal time base the wavefo...

Страница 242: ...USB storage device when detected in CSV format For detailed saving operation refer to descriptions in Store and Load Jump to Set the operating status of the oscilloscope to STOP Then select the data o...

Страница 243: ...apply them to I2S decoding function set the corresponding I2S decoding parameters automatically For analog channels the threshold level settings will also be copied 3 Source setting Press Sources to e...

Страница 244: ...it The available channels include CH1 CH4 and D0 D15 When you select the analog channel CH1 CH4 press Data Thre below Data then rotate the multifunction knob or use the pop up numeric keypad to set t...

Страница 245: ...le or disable the label display of the I2S decoding bus When enabled the bus label I2S will be displayed at the upper left side of the bus when the bus display is enabled Event table Press Event Table...

Страница 246: ...clude Hex Dec Bin and ASC Set the view type Press View then rotate the multifunction knob to select the form of the event table to be viewed Press down the knob to select it You can also press View co...

Страница 247: ...oding Option The oscilloscope samples the 1553B signal and judges each data point to be logic 1 or logic 0 according to the preset threshold level 1553B decoding is required to specify the data channe...

Страница 248: ...bus When enabled the bus label 1553B will be displayed at the upper left side of the bus when the bus display is enabled Event table Press Event Table to enter the event table setting menu The event...

Страница 249: ...t the view type Press View then rotate the multifunction knob to select the form of the event table to be viewed Press down the knob to select it You can also press View continuously or enable the tou...

Страница 250: ...ding event tables of four decode modules Decode1 Decode2 Decode3 and Decode4 Press Decode and rotate the multifunction knob to select a decode module then press down the knob to select it You can also...

Страница 251: ...e Contents in this chapter To Enable Ref Function To Select the Reference Channel To Select the Ref Source To Adjust the Ref Waveform Display To Save to Internal Memory To Clear the Display of the Ref...

Страница 252: ...select the desired reference waveform channel Ref1 Ref10 You can also press Current continuously or enable the touch screen to select the reference waveform channel By default Ref1 is enabled To Sele...

Страница 253: ...r the display of the current reference waveform on the screen Meanwhile the VScale and VOffset menus for the reference waveform are grayed out and disabled You can also press CLEAR on the front panel...

Страница 254: ...etting Press More Label to enter the label setting menu You can use the built in label in the library or manually input a label Three input methods are available for you to input a label name includin...

Страница 255: ...rt function is valid Note MSO5000 series oscilloscope only supports the flash memory USB storage device of FAT32 format To Import from Internal or External Memory You can also import the stored refere...

Страница 256: ...www GlobalTestSupply com nd Quality Products Online at sales GlobalTestSupply co...

Страница 257: ...tion of MSO5000 series oscilloscope can accomplish this task perfectly Contents in this chapter To Enable or Disable the Pass Fail Test To Start or Stop the Pass Fail Test Operation To Select the Sour...

Страница 258: ...e test mask the display status of the test information and the output form of a failed test based on the selected source channel for the test Then save the test mask range to the internal or external...

Страница 259: ...at the right Press CursorA and CursorB and rotate the multifunction knob to adjust the positions of both Cursor A and Cursor B respectively Also you can press CursorAB and rotate the multifunction kno...

Страница 260: ...e width Press Polarity continuously to select Positive or Negative Then press Pulse rotate the multifunction knob or use the pop up numeric keypad to set the pulse width Its range is from 100 ns to 10...

Страница 261: ...able the touch screen to tap the icon at the right upper corner of the statistics window to close it The test results statistics include the number of failed frames the number of successful frames and...

Страница 262: ...www GlobalTestSupply com nd Quality Products Online at sales GlobalTestSupply co...

Страница 263: ...corded waveforms of the analog input channels CH1 CH4 and the digital channels D0 D15 enabling you to analyze the waveforms better Note The horizontal time base must be set to YT mode during waveform...

Страница 264: ...f frames for recording The time difference T at the upper right corner indicates the time different between the current frame and the first frame After the recording is completed automatically turns o...

Страница 265: ...use the pop up numeric keypad to set the time interval between frames The available range is from 10 ns to 10 s 2 Recording Frames The recording frames refer to the number of frames that can be record...

Страница 266: ...set the playback mode to cycle or single plays from the start frame to the end frame then such playback operation is repeated until you stop it manually plays from the start frame to the end frame and...

Страница 267: ...End Frame Press End Frame then rotate the multifunction knob or use the pop up numeric keypad to set the end frame of playback The default is the number of frames of the recorded waveforms www Global...

Страница 268: ...www GlobalTestSupply com nd Quality Products Online at sales GlobalTestSupply co...

Страница 269: ...ion allows you to search for relevant events based on the search condition that you set The navigation function includes the recording playing navigation time navigation and event navigation Contents...

Страница 270: ...ype For Pulse Set menu settings refer to descriptions in Pulse Trigger Press Threshold to enter the threshold setting menu Rotate the multifunction knob or use the pop up numeric keypad to set the thr...

Страница 271: ...search type to the search settings For example if the current trigger type is Edge Trigger press From Trigger to copy from the Edge trigger settings to the Edge search settings Note If you select From...

Страница 272: ...e navigation combination keys to play the recorded waveforms Press to play the recorded waveforms in reverse order press to play the recorded waveforms in normal order press to stop playing Time Navig...

Страница 273: ...er s Guide 15 5 number in the mark table decreases press to navigate to the next event the serial number in the mark table increases the key is disabled in the event navigation www GlobalTestSupply co...

Страница 274: ...www GlobalTestSupply com nd Quality Products Online at sales GlobalTestSupply co...

Страница 275: ...orm display type persistence time intensity grid type grid brightness and etc Contents in this chapter To Select the Display Type To Set the Persistence Time To Set the Waveform Intensity To Set the S...

Страница 276: ...to select it You can also press Persis Time continuously or enable the touch screen to select it The available values are Min 100 ms 200 ms 500 ms 1 s 2 s 5 s 10 s and Infinite In the following part...

Страница 277: ...select it You can also press Grid continuously or enable the touch screen to select it You can select FULL HALF NONE or IRE FULL turns the background grid and coordinate on HALF turns the background...

Страница 278: ...are displayed on the screen to indicate the times of data acquisition or acquisition probability Waveform Freeze Press More Waveform Freeze to enable or disable waveform freeze function If enabled th...

Страница 279: ...n the filename input interface is displayed e g inputting RIGOL Scope For details refer to descriptions in To Create a Folder 2 Set the Wait Time Press Time to Start rotate the multifunction knob or u...

Страница 280: ...www GlobalTestSupply com nd Quality Products Online at sales GlobalTestSupply co...

Страница 281: ...t convenience for engineers who need to use the signal source and oscilloscope at the same time This chapter introduces how to use the built in signal source As the functions and setting methods of th...

Страница 282: ...ad to set the frequency or period value of the current signal Different waveforms have different frequency or period reciprocal of the frequency ranges Sine 100 mHz to 25 MHz Square 100 mHz to 15 MHz...

Страница 283: ...o channels and stably display the waveforms After switching the channel status the phase deviation between the two waveforms is changed At this time press Align Phase and then the phase deviation show...

Страница 284: ...mmetry Definition Press Settings Symmetry then rotate the multifunction knob or use the pop up numeric keypad to set the symmetry of the ramp waveform The settable range is from 0 to 100 To Output Pul...

Страница 285: ...to select Noise Press down the knob to select it You can also press Wave continuously or enable the touch screen to select it At this time you can set the amplitude high level offset low level and imp...

Страница 286: ...ble the touch screen to select it You can set the parameters for the Sinc waveform as shown in the figure below For setting methods refer to descriptions in To Output Sine Figure 17 3 Sinc Exp Rise Pr...

Страница 287: ...tting methods refer to descriptions in To Output Sine Figure 17 5 Exp Fall ECG Press Wave rotate the multifunction knob to select ECG Press down the knob to select it You can also press Wave continuou...

Страница 288: ...also press Wave continuously or enable the touch screen to select it At this time you can set the parameters for the Lorentz waveform as shown in the figure below For setting methods refer to descript...

Страница 289: ...Load from CH to enter the channel loading menu You can set the channel signal required to be loaded Channel selection Press Channel and rotate the multifunction knob to select the desired channel CH1...

Страница 290: ...e or High Level continuously to select it Also you can enable the touch screen to select it Rotate the multifunction knob or use the pop up numeric keypad to set the amplitude or high level value of t...

Страница 291: ...e voltage value of the current point Its available range is from 2 5 V to 2 5 V Insert Press Insert to insert a new editable point in the middle position between the current editing point and the next...

Страница 292: ...f you select High Level for this menu then the menu below the High Level is automatically turns out to be Low Level Set the offset and low level First select Offset or Low Level then rotate the multif...

Страница 293: ...se refer to descriptions in Store and Load to save the currently edited waveform file to the internal or external memory in arb format you can overwrite the original file or save the currently edited...

Страница 294: ...ob or use the pop up numeric keypad to set the frequency of the modulation signal The available range is from 1 Hz to 50 kHz 3 Set the Modulation Depth The modulation depth refers to the strength of t...

Страница 295: ...nob or use the pop up numeric keypad to set the frequency of the modulation signal The available range is from 1 Hz to 50 kHz 3 Set the Frequency Offset Press Deviation then rotate the multifunction k...

Страница 296: ...orm Generator of MSO5000 series oscilloscope supports the sweep function In the sweep mode the Function Arbitrary Waveform Generator outputs from the start frequency to the end frequency at a specifie...

Страница 297: ...ic keypad to set the end frequency By default the start frequency is 100 Hz and the end frequency is 1 kHz The ranges of start frequency and end frequency for different sweep waveforms are different A...

Страница 298: ...ime then rotate the multifunction knob or use the pop up numeric keypad to set the sweep time By default it is 1 s Its available range is from 1 ms to 500 s After modifying the sweep time the Function...

Страница 299: ...te the multifunction knob or use the pop up numeric keypad to set the desired delay time The range of the delay time related to the burst cycles 4 Burst Setting Press Settings to enter the burst setti...

Страница 300: ...ser s Guide Press Burst Period then rotate the multifunction knob or use the pop up numeric keypad to set the desired burst period Its available range is from 1 ms to 500 s By default it is 10 ms www...

Страница 301: ...when needed You can also copy a new file of the specified type as well as delete and rename the specified type of file in the internal memory or external USB storage device via the disk management men...

Страница 302: ...be saved For filename input refer to descriptions in To Create a Folder Press More Auto Name to select ON then the filename is generated automatically If the filename that you input is the same as the...

Страница 303: ...then the available storage channels include CH1 CH4 and D0 D15 Only when you select Memory under Data Source and the format is bin or cvs can you set the storage channel Press Window to display or hi...

Страница 304: ...face is displayed automatically By default Local Disk C is selected 3 Rotate the multifunction knob or enable the touch screen to open the internal memory storage interface 4 Rotate the multifunction...

Страница 305: ...orage interface 3 Rotate the multifunction knob to select the desired storage location and then press down the knob to open it You can also enable the touch screen to select the storage location Rotat...

Страница 306: ...ing refer to descriptions of Wave in Storage Type 5 Press Save to save the waveform file to the selected folder in the specified save format 2 Load the Specified Type of File from the External USB Sto...

Страница 307: ...Press Storage Disk File Type rotate the multifunction knob to select the desired file type The default is In the current path only the file whose suffix of its file name matches that of the file type...

Страница 308: ...put Method For example create a folder with the name Filename Tip During the name input rotate the multifunction knob to select the desired contents Press down the knob to input the desired contents o...

Страница 309: ...in the current page press to go to the next page until you find the desired character 4 Input the Other Characters Use the same method to input the remaining letters ilename by referring to Step 3 Pa...

Страница 310: ...no character in the Pinyin Input Area and Name Input Area please go to Step 3 If there are characters in the Name Input Area and Pinyin Input Area press rotate the multifunction knob to select the de...

Страница 311: ...ting from the last one to the character to be modified then input the new character again If the touch screen is enabled tap to directly move the cursor to the character required to be modified or del...

Страница 312: ...nal or external memory Press down the knob to open it 2 Rotate the multifunction knob to select the file or folder to be renamed 3 Press Rename and then the filename input interface is displayed For t...

Страница 313: ...Restore default is displayed Press OK or tap OK to restore the oscilloscope to the factory default settings as shown in the table below Table 18 1 Factory Settings Parameter Factory Settings Horizonta...

Страница 314: ...nti Aliasing Off Trigger Trigger Type Edge Trigger Source Selection CH1 Edge Type Rising Trigger Mode Auto Trigger Coupling DC Trigger Holdoff 16 ns Noise Rejection Off Display Display Type Vector Per...

Страница 315: ...e CH1 AY 100 mV BY 100 mV Vertical Unit Source Hori Unit s Region Main Track AX Source CH1 BX Source CH1 AX 1 s BX 1 s Track X XY AX 233 3 mV BX 100 mV AY 233 3 mV BY 100 mV Measure Indicator Off Stor...

Страница 316: ...1 Operate Stop Range Screen X Mask 0 24 div Y Mask 0 48 div File Type pf File Type pf File Name RigolDS0 Window Show Indicator OFF Aux Output OFF Output Event Fail Polarity Positive Pulse 1 s Err Acti...

Страница 317: ...to Config Peak to Peak OFF CH All Overlay OFF Coupling OFF Printer Copies 1 Paper Size A4 Ink Saver ON Printer HP Laserjet Email Attachment Screen Quick Settings Operation Save Image Format png Invert...

Страница 318: ...T Operation OFF Source CH1 Offset 0 dB Scale 1 dB Unit dB dBV X Start End Start Freq 0 Hz Stop Freq 10 MHz Window Function Hanning Mode Trace View Full Peak Search OFF Peak Number 5 Threshold 5 5 dB E...

Страница 319: ...ize Medium Thre CH1 0 V Thre CH2 0 V Thre CH3 0 V Thre CH4 0 V Sensitivity 0 3 Div A Operation OFF SourceA CH1 Wave Size Medium Thre CH1 0 V Thre CH2 0 V Thre CH3 0 V Thre CH4 0 V Sensitivity 0 3 Div...

Страница 320: ...Offset 0 U Scale 2 U Invert OFF Exp Operation OFF Source CH1 Offset 0 U Scale 1 U Invert OFF Abs Operation OFF Source CH1 Offset 0 U Scale 20 mV Invert OFF Low Pass Operation OFF Source CH1 Offset 0...

Страница 321: ...ert OFF AX B Operation OFF Source CH1 Offset 0 V Scale 20 mV A 0 B 0 Invert OFF Logic Analyzer LA Select None D7 D0 OFF D15 D8 OFF Wave Size Medium Arrange D15 D0 Threshold D7 D0 1 4 V D15 D8 1 4 V La...

Страница 322: ...I2C Bus Status OFF Clock CH1 SCL Thre 0 V SDA CH2 SDA Thre 0 V Exchange SCL SDA R W Without SPI Bus Status OFF Mode Timeout Timeout 1 s CLK CH1 Threshold 0 V Slope Rising MISO CH2 Threshold 0 V MOSI...

Страница 323: ...le Position 50 I2S Bus Status OFF SCLK CH1 SCLK Thre 0 V SCLK Edge Rising WS CH2 SCLK Thre 0 V Data CH3 Data Thre 0 V Word Size 4 Receive 4 Alignment I2S WS Low Left Endian MSB Data Polarity 1553B Bus...

Страница 324: ...00 User s Guide Note 1 It is only available for the model installed with the MSO5000 AWG option Note 2 PLA2216 active logic probe option is required to be ordered www GlobalTestSupply com nd Quality P...

Страница 325: ...lity Function Setting In the system function setting menu you can set the parameters for the remote interface and the system related parameters Contents in this chapter Remote Interface Configuration...

Страница 326: ...network The network interface of the oscilloscope is on the rear panel Press Utility IO LAN to open the LAN Settings interface as shown in the following figure You can view the current network settin...

Страница 327: ...IP address ranging from 169 254 0 1 to 169 254 255 254 and the subnet mask 255 255 0 0 automatically according to the current network configuration Only when DHCP is not selected or failed to be conn...

Страница 328: ...n The range of the first segment nnn is from 0 to 223 except 127 and the range for the other three segments is from 0 to 255 You are recommended to ask your network administrator for a gate address av...

Страница 329: ...status and parameters of the HDMI video output interface on the rear panel Note that only when the HDMI interface is connected can this menu be enabled Press HDMI to enter the HDMI setting menu Press...

Страница 330: ...he GPIB cable to connect the instrument to the PC Press Utility IO GPIB then rotate the multifunction knob directly to set the GPIB address or use the pop up numeric keypad to input the GPIB address I...

Страница 331: ...en press down the knob to select the language You can also press Language continuously to select the language or enable the touch screen to tap the desired language and select it System Information Pr...

Страница 332: ...e Aux output the AUX Out menu is automatically set to PassFail When the AUX Out menu is automatically set to TrigOut then in the pass fail test menu the Aux Out function is automatically disabled Help...

Страница 333: ...figure Figure 19 2 Self calibration Interface Press Start and then the oscilloscope will start to execute the self calibration program After the self calibration program is launched the Exit menu is...

Страница 334: ...e knob to select it You can also press Printer continuously or enable the touch screen to select it Set the IP Address Press Printer IP then rotate the multifunction knob to set each segment of the IP...

Страница 335: ...a emails Press Utility More Email to enter the email setting menu 1 Set the Email Server Press Settings to enter the email server login setting menu The following window is displayed After the setting...

Страница 336: ...ed input methods of the receiver email address refer to descriptions in To Create a Folder 3 Upload Attachment Press Attachment then rotate the multifunction knob to select the attachment type to be u...

Страница 337: ...ously to select among the options or enable the touch screen to tap the desired operation type Image Format Press Format and then rotate the multifunction knob to select the image format You can also...

Страница 338: ...elect Save Setup Press down the knob to select it You can also press Operation continuously to select among the options or enable the touch screen to tap the desired operation type You do not need to...

Страница 339: ...press Operation continuously to select among the options or enable the touch screen to tap the desired operation type 7 Email When the Quick shortcut key is set to Email press Quick on the front panel...

Страница 340: ...d test interface virtual keypad of the front panel At this time you can press the keys on the front panel to check whether the virtual keys are highlighted If yes it indicates that the keys work norma...

Страница 341: ...ower right side of the screen in hh mm hour minute format When you save the waveform the output file will contain the time information Press Utility More Time to open the time setting menu 1 Set the S...

Страница 342: ...6 Set Minute Press Minute rotate the multifunction knob directly to set it or use the pop up numeric keypad to input the numeric value to set it The setting range of each item of the system time conf...

Страница 343: ...MSO5000 series digital oscilloscope supports Web Control When the instrument is connected to the internet Web Control can display the waveform interface of the oscilloscope in a real time manner You...

Страница 344: ...is a USB TMC device After you connect the oscilloscope to the PC properly and power it on for the first time the oscilloscope will be automatically configured with the USB interface In this case Found...

Страница 345: ...Chapter 20 Remote Control RIGOL MSO5000 User s Guide 20 3 5 6 3 4 www GlobalTestSupply com nd Quality Products Online at sales GlobalTestSupply co...

Страница 346: ...y connected to the PC via the USB interface You can also click to search for the resource 4 View the device resource The searched instrument resource will be under the directory of RIGOL Online Resour...

Страница 347: ...ly Right click the resource name MSO5354 USB0 0x1AB1 0x3514 MSO53541234567890 INSTR and select SCPI Panel Control to turn on the remote command control panel through which you can send commands and re...

Страница 348: ...ne Resource directory For example MSO5354 TCPIP 172 16 3 14 INSTR 5 Control the instrument remotely Right click the resource name MSO5354 TCPIP 172 16 3 14 INSTR and select SCPI Panel Control to turn...

Страница 349: ...the GPIB Address Set the GPIB address of the oscilloscope according to the description in To Set the GPIB Address 4 Search for the device resource Start Ultra Sigma and then click A window is display...

Страница 350: ...rmally if not please follow the corresponding prompt messages to handle it 5 View the device resource The searched instrument resource will be under the directory of RIGOL Online Resource as shown in...

Страница 351: ...there are signals generated from the item to be tested you can connect the probe compensation output signal to the faulty channel to locate the problem and then determine whether the channel or the i...

Страница 352: ...2 Check whether the screen or your finger is stained with oil or sweat If yes please clean the screen or dry your hands 3 Check whether there is a strong magnetic field around the instrument If the i...

Страница 353: ...FPC Rack Mount Kit MSO5000 RM USB GPIB Interface Converter USB GPIB Near field Probe NFP 3 Power Analysis Phase Difference Correction Jig RPA246 Digital Oscilloscope Demonstration Plate DK DS6000 Ban...

Страница 354: ...O5000 AUTO FlexRay Serial Bus Trigger and Analysis FlexRay MSO5000 FLEX Audio Serial Bus Trigger and Analysis I2S MSO5000 AUDIO MIL STD 1553 Serial Bus Trigger and Analysis MIL STD 1553 MSO5000 AERO M...

Страница 355: ...for the defective product To get repair service please contact with your nearest RIGOL sales or service office There is no other warranty expressed or implied except such as is expressly set forth her...

Страница 356: ...www GlobalTestSupply com nd Quality Products Online at sales GlobalTestSupply co...

Страница 357: ...28 17 2 Gauss 17 8 GPIB 19 6 Hamming 6 9 Hanning 6 9 Haversine 17 9 Horizontal Expansion 4 10 Horizontal Position 3 3 I2C Decoding 11 14 I2C Trigger 5 34 I2S Trigger 5 45 Idle Time 5 31 Impedance 17...

Страница 358: ...17 18 Sweep Trigger Source 17 18 17 19 Symmetry 17 4 Sync 5 15 To Create the Waveform 17 10 Triangle 6 9 Trigger Coupling 5 5 Trigger Holdoff 5 6 Trigger LEVEL Threshold Level 5 3 Trigger Mode 5 4 Tri...

Отзывы: