
−
12
−
ME12044A
6.5.3
Setting the detector provisionally
The I
0
-detector (proportional counter) is used to adjust the monochromator crystal. Initialize the
I
0
-detector. Select Adjust
→
I
0
Counter, enter HV = 1050, Base line = 0.5, Window = 3, and click
on [OK].
Figure 6.3 Detector Setup Screen
6.5.4
Setting the slits
Set the divergence slit (DS) and receiving slit (RS). Set the DS to 2 mm and the RS to 0.2 mm.
Figure 6.4 Slit Setup Screen