20
13
b) With a frequency range over 150 kHz to 80 MHz, the field strengths should be less than 3 V/m.
Recommended distances between portable and mobile RF communications equipment
and the ri-mini® and pen-scope®
The ri-mini® and pen-scope® are intended for use in an electromagnetic environment in which
RF emissions are controlled. The customer or user of the ri-mini® and pen-scope® can help to
avoid electromagnetic interference by observing the minimum distance between portable and
mobile RF communications equipment (transmitters) and the ri-mini® and pen-scope® in
accordance with the maximum output power of the communication equipment.
Rated maximum
output power of the
transmitter
(W)
Separation distance according to the frequency of the transmitter
(m)
150 KHz to 80 MHz
80 MHz to 800 MHz
800 MHz to 2.7 GHz
0.01
0.12
0.12
0.23
0.23
0.1
0.38
0.38
0.73
0.73
1
1.2
1.2
2.3
2.3
10
3.8
3.8
7.3
7.3
100
12
12
23
23
For transmitters with a maximum output power not listed above, the recommended distance d
in metres (m) can be estimated using the equation for the transmitter frequency, where P is the
maximum output power of the transmitter in watts (W). according to the transmitter
manufacturer.
NOTE 1 At 80 MHz and 800 MHz, the separation distance applies to the higher frequency
range.
NOTE 2 These guidelines may not apply in all situations. The electromagnetic propagation is
affected by absorption and reflection of structures, objects and people.
Содержание pen-scope
Страница 39: ...39...
Страница 50: ...50 F II 2 P 2 1 H Riester 2 2 alkaline AA 1 5 IEC LR6 2 3 2 4 2 5 F...
Страница 62: ...62 1 Riester 2 Rudolf Riester GmbH 4 Dept Repairs RR 2 Bruckstr 31 72417 Jungingen Germany...
Страница 63: ...63...