Guidelines and manufacturer's declaration – electromagnetic immunity
The ri-scope® L/EliteVue instruments are intended for use in the electromagnetic environ-
ment specified below. The customer or user of the ri-scope® L/EliteVue should ensure that it
is used in such an environment.
Immunity
testing
IEC 60601
Testlevel
Compli-
ance
Electromagnetic environment -
guidance
IEC 61000-4-6
Conducted RF
disturbances
according to IEC
61000-4-6
3 Vrms
0.5 MHz to 80 MHz
6 V in ISM frequen-
cy bands
Between 0.15 MHz
and 80 MHz
80% AM at 1 kHz
Not
applicable
Portable and mobile RF communica-
tions equipment should not be used
closer to any part of the non-contact
ri-pen, including the cables, than the
recommended distance, which is cal-
culated using the equation applicable
to the transmitter frequency.
Recommended separation distance:
d = 1.2 x P 80 MHz to 800 MHz
d = 2.3 x P 800 MHz to 2.7 GHz
Where P is the maximum output po-
wer of the transmitter in watts (W)
according to the transmitter manu-
facturer and d is the recommended
distance in metres (m).
Field strengths of fixed RF transmit-
ters, determined by an electroma-
gnetic site survey, should be lower
than the compliance standard in each
frequency range.
Interference may occur in the vicinity
of devices marked with the following
symbol:
Emitted RF
IEC 61000-4-3
Proximity fields
of wireless RF
communications
equipment
3 V/m
80 MHz bis 2,7 GHz
380 – 390 MHz
27 V/m, PM 50%,
18 Hz
430 – 470 MHZ
28 V/m, (FM ± 5
kHz, 1 kHz sine)
PM, 18 Hz11
704 – 787 MHz
9 V/m, PM 50%,
217 Hz
800 – 960 MHz
28 V/m, PM 50%,
18 Hz
1700 – 1990 MHz
28 V/m, PM 50%,
217 Hz
2400-2570 MHz
28 V/m, PM 50%,
217 Hz
5100-5800 MHz
9 V/m, PM 50%,
217 Hz
10 V/m
27 V/m
28 V/m
9 V/m
28 V/m
28 V/m
9 V/m
34