RHEONIK Messgeräte GmbH
REV. I, Apr. 2004
.
Operation Manual RHE 07, 08, 11
Page
16
Note:
With the ordered option density measurement, you can select volumetric rather than gravimetric
units. Please use the + or – keys to choose the appropriate setup for each parameter.
–
. Standard
units for each parameter are:
SI
ANSI
_____________________________________________________________________
Totalizer:
t, kg, g,
tn, lb, oz
Mass flow :
t, kg, g / h, min, sec
tn, lb, oz / h, min, sec
Density:
kg/l (= kg/dm3)
lb/ga, BaumeL, BaumeH, kg/l @ *)
Temperature:
°C
°F
Totalizer volume:
m3, l, ml
ga, ba, in3
* This density unit is referenced to a specific temperature
(density at reference temperature).
The selection may be changed as often as desired and will be held in non volatile
EEProm
memory.
Only certain units can be selected, depending on the type of sensor chosen.
2.1.3.4. Display sequence and Format
In order to determine the sequence of different measurement data displays you have to program first,
second , .. display function (
1.Disp = XXXXX
).
In
TOGGLE
mode the LCD display switches every 10 seconds to one of the two possible measure-
ment data displays (
Disp = Toggle
).
Totalizer increment resolution is selected in total format display (
TotalForm = X.XX
)
.
Consider the maximum totalizer of 8 digits! Totalizer overflow will be indicated.
With
Show Errors = off
no error messages will be indicated on display.
After
Lock Keys = on
the keyboard will be blocked until next power OFF and ON.