NOTES FOR INTRINSICALLY SAFE INSTALLATION
Only equipment with safety agency labels attached to the sensor and transmitter meet the
agency approval requirements.
Intrinsically safe flow meters must be installed according to the wiring diagram, supplied with
the meter.
Observe proper earth ground wiring according to this diagram.
Sensor cable must be suitable for the sensor operating temperature range.
All intrinsically safe cables must be installed separated from all other cables.
Observe temperature class and maximum allowable sensor temperature, indicated on sensor
type label, for safe operation.
All electrical installations must comply with national and local codes.
1.4. Installation instructions for the RHE flow transmitter
Install the RHE unit in an area where the ambient temperature falls within the range
-40°C .. +60°C. For installations outside this range please consult the manufacturer.
Locations with extreme vibrations must be avoided.
Do not locate the flow transmitter in direct sunlight.
Sensor RHM and transmitter RHE were calibrated together at the factory.
Therefore make sure that the serial numbers of connected systems comply with
serial numbers indicated on instrument type labels.
1.5. Power supply wiring
The remote electronic unit RHE is delivered set up for 230 VAC, 115 VAC or 24 VDC power input.
The power supply must be turned off while wiring to the remote unit RHE ...
Power supply v-10% must correspond to the voltage indicated on electronic type label or in
the power supply wiring compartment.
The earth grounding of the power supply must be connected to the RHE power supply wiring section.
Failure to connect earth ground will nullify the intrinsic safety !