TROUBLESHOOTING - NO UV LIGHT
E-49
Check UV lamp
Visually check the UV bulb. If it
is broken or damaged replace it.
Make sure the UV lamp is in
place correctly.
Check UV lamp power
cord and ballast
Check main circuit board
NO UV LIGHT
If the UV lamp cord or ballast is
broken, replace it.
If the UV lamp cord or ballast is
broken, replace it.
Step 1
Step 2
Step 3
OK
OK
ENGLISH
TROUBLESHOOTING CHART
Repair tips:
1. Check UV lamp
2. Check UV lamp power cord and ballast
3. Check main circuit board
Some of these steps expose dangerous
high voltage. Only qualified service
technician should attempt this procedure.
WARNING