Modbus
MI3
91
Start
address
Size
[bits]
Modbus Access Data
Type
Content
Values
MI3[M]
command
100
8
discretes input bit field Get connected heads bit 0: head 1 ..
bit 7: head 8
bit high: head connected
bit low: head disconnected
HC
110
8
discretes input bit field Get registered heads bit 0: head 1 ..
bit 7: head 8
bit high: head registered
bit low: head not registered
HCR
120
16
holding register short
Laser control (only with
laser)
0: off, 1: on
XL
130
16
holding register short
Relay alarm output
control
0: off
1: on
2: norm. open
3: norm. closed
KB
410
32
input register
float
analog input 1 value
0 .. 5 V
TV1I
420
32
input register
float
analog input 2 value
0 .. 5 V
TV2I
430
16
input register
short
Trigger
0: off
1: on
XT
5<k>0 16
holding register short
analog output k mode
5: TCJ
6: TCK
7: TCR
8: TCS
9: 0...5 V
10: 0...10 V
99: disable (tristate)
XO<k>O
5<k>1 16
holding register short
analog output k source head number or 0:
fixed value from 5<k>3
O<k>O
5<k>2 16
holding register short
analog output k source
parameter
1: internal temp. of 5<k>1
2: object temp. of 5<k>1
O<k>O
5<k>3 32
holding register float
analog output k fixed
temp. value
value within range set in address 5<k>0 -
5<k>5 32
holding register float
analog output k bottom
temp. value
device bottom temp. .. device top temp. L<k>O
5<k>7 32
holding register float
analog output k top
temp. value
device bottom temp. .. device top temp. H<k>O
1.800.868.7495
Содержание MI3
Страница 2: ...1 800 868 7495 info Raytek Direct ca...
Страница 9: ...20 NOTES 119 1 800 868 7495 info Raytek Direct ca...
Страница 73: ...Accessories MI3 Rev F1 05 2013 73 Figure 66 Protective Window 1 800 868 7495 info Raytek Direct ca...
Страница 113: ...ASCII Programming MI3 Rev F1 05 2013 113 1 800 868 7495 info Raytek Direct ca...
Страница 119: ...Notes MI3 Rev F1 05 2013 119 20 Notes 1 800 868 7495 info Raytek Direct ca...