ASCII Programming
MI3
105
If the ambient temperature is not requested by the internal head temperature, you must set the
ambient temperature values, as follows:
A=250.0
Ambient temperature (example)
AC=1
Control ambient background temp. compensation
18.4.2
Temperature Pre-Processing
The samples from the AD converter (energy values) can be processed before temperature calculation.
The corresponding filter command is <FF>.
18.4.3
Emissivity Setting and Alarm Set points
The device allows three choices for the emissivity setting and two for the alarm output setting.
ES
Selection of the emissivity setting.
ES=1
Emissivity set by a constant number according to the „E“ command
ES=E
Emissivity set by a voltage on FTC1 (analog input)
ES=D
Emissivity set by the entries in a table (selected by digital inputs FTC1 – FTC3)
?CE
asks for the emissivity value that is actually used for temperature calculation
There are eight entries possible for emissivity setting
①
and a related set point (threshold)
②
. To be
able to write or read these values, use the following commands:
EP=2
set pointer for table entry, e.g. to line 2
③
EV=0.600
set the emissivity value for line 2 to 0.600
④
SV=220.0
set the set point (threshold) for line 2 to 220.0
⑤
Figure 83: Table for Emissivity and Set Points
To activate these emissivity settings, you need to have the 3 external inputs (FTC) connected.
According to the digital combination on the FTC wires, one of the table entries will be activated, see
section 7.2
①
②
③
④
⑤
1.800.868.7495
Содержание MI3
Страница 2: ...1 800 868 7495 info Raytek Direct ca...
Страница 9: ...20 NOTES 119 1 800 868 7495 info Raytek Direct ca...
Страница 73: ...Accessories MI3 Rev F1 05 2013 73 Figure 66 Protective Window 1 800 868 7495 info Raytek Direct ca...
Страница 113: ...ASCII Programming MI3 Rev F1 05 2013 113 1 800 868 7495 info Raytek Direct ca...
Страница 119: ...Notes MI3 Rev F1 05 2013 119 20 Notes 1 800 868 7495 info Raytek Direct ca...