Operation
MI3
49
<OUT1 Mode>
defines the mode for the analog output:
<TCJ>, <TCK>, <TCR>, <TCS>
<0-5V>
<0-10V>
<disable> output goes to high-resistance
<OUT1 Source>
assigns the selected head to the analog output:
<#1>, <#2>, …, <Head
max
>
<OUT1 Value>
defines the basis for the output value:
<Tobject>: object temperature to be output
<Tambient>: head ambient temperature to be output
<OUT1 low temp.>
defines the temperature for the low end of the analog output range (scaling)
<OUT1 high temp.>
defines the temperature for the high end of the analog output range (scaling)
<OUT2 Mode>
defines the mode for the analog output 2:
<0-20mA>
<4-20mA>
<0-5V>
<0-10V>
<disable> output goes to high-resistance
<OUT2 Source>
assigns the selected head to the analog output:
<#1>, <#2>, …, <Head
max
>
<OUT2 Value>
defines the basis for the output value:
<Tobject>: object temperature to be output
<Tambient>: head ambient temperature to be output
<OUT2 low temp.>
defines the temperature for the low end of the analog output range (scaling)
<OUT2 high temp.>
defines the temperature for the high end of the analog output range (scaling)
<Interface>
RS485/Profibus/Modbus:
<address>: unique address of the box in the network.
<baudrate>: baud rate for the box. Each device in the network must be set to the
same baud rate. The baud rate for Profibus is automatically negotiated between
master and slave.
Profinet/Ethernet:
<DHCP>: on/off – network protocol to configure a device in a network. DHCP
is switched off for Profinet communications.
<IP address>: a unique address of the box in the network, only changeable at
DHCP = off
<SubNetMask>: the subnet mask defines the interpretation of the IP address,
only changeable at DHCP = off
<Gateway>: a gateway connects two subnets at different subnet addresses, only
changeable at DHCP = o
ff
<Port>: communication
port, only changeable at DHCP = off,
read-onlyfor
Profinet communications
<MAC>: MAC address, read-only
<Factory default>
sets the box back to factory default.
The factory default values are to be found in section 18.9
page 108.
1.800.868.7495
Содержание MI3
Страница 2: ...1 800 868 7495 info Raytek Direct ca...
Страница 9: ...20 NOTES 119 1 800 868 7495 info Raytek Direct ca...
Страница 73: ...Accessories MI3 Rev F1 05 2013 73 Figure 66 Protective Window 1 800 868 7495 info Raytek Direct ca...
Страница 113: ...ASCII Programming MI3 Rev F1 05 2013 113 1 800 868 7495 info Raytek Direct ca...
Страница 119: ...Notes MI3 Rev F1 05 2013 119 20 Notes 1 800 868 7495 info Raytek Direct ca...